Reproducibility of image quality for moving objects using respiratory-gated computed tomography: A study using a phantom model

机译:使用呼吸门计算机断层摄影术对运动物体的图像质量的再现性:使用幻像模型的研究

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

To investigate the reproducibility of computed tomography (CT) imaging quality in respiratory-gated radiation treatment planning is essential in radiotherapy of movable tumors. Seven series of regular and six series of irregular respiratory motions were performed using a thorax dynamic phantom. For the regular respiratory motions, the respiratory cycle was changed from 2.5 to 4 s and the amplitude was changed from 4 to 10 mm. For the irregular respiratory motions, a cycle of 2.5 to 4 or an amplitude of 4 to 10 mm was added to the base data (i.e. 3.5-s cycle, 6-mm amplitude) every three cycles. Images of the object were acquired six times using respiratory-gated data acquisition. The volume of the object was calculated and the reproducibility of the volume was decided based on the variety. The registered image of the object was added and the reproducibility of the shape was decided based on the degree of overlap of objects. The variety in the volumes and shapes differed significantly as the respiratory cycle changed according to regular respiratory motions. In irregular respiratory motion, shape reproducibility was further inferior, and the percentage of overlap among the six images was 35.26 in the 2.5- and 3.5-s cycle mixed group. Amplitude changes did not produce significant differences in the variety of the volumes and shapes. Respiratory cycle changes reduced the reproducibility of the image quality in respiratory-gated CT.
机译:在呼吸门控放射治疗计划中研究计算机断层扫描(CT)成像质量的可重复性对于活动性肿瘤的放射治疗至关重要。使用胸部动态体模进行了七个系列的常规呼吸运动和六个系列的不规则呼吸运动。对于常规呼吸运动,呼吸周期从2.5 s更改为4 s,幅度从4 mm更改为10 mm。对于不规则的呼吸运动,每三个周期将2.5到4的周期或4到10 mm的振幅添加到基本数据中(即3.5 s周期,6 mm振幅)。使用呼吸门控数据采集对物体的图像采集六次。计算对象的体积,并根据品种决定体积的可重复性。添加物体的配准图像,并根据物体的重叠程度确定形状的可复制性。体积和形状的变化随着呼吸周期根据常规呼吸运动而变化而显着不同。在不规则呼吸运动中,形状再现性更差,在2.5和3.5 s循环混合组中,六个图像之间的重叠百分比为35.26。振幅变化在体积和形状的变化上没有产生显着差异。呼吸循环变化降低了呼吸门控CT图像质量的可重复性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号