Gafchromic film has been used for measurement of computed tomography (CT) dose distributions within phantoms. The film was calibrated in the beam from a superficial therapy unit and the accuracy confirmed by comparison with measurements with a 20mm long ionisation chamber. The results have been used to investigate approaches to CT dosimetry. Dose profiles were recorded within standard CT head and body phantoms and scatter tail data fitted to exponential functions and extrapolated to predict dose levels in longer phantoms. The data have been used to simulate both CT dose index (CTDI) measurements with ionisation chambers of differing length and measurements of cumulative doses with a 20mm chamber for scans of varying length. The results show that the length of a pencil ionisation chamber is the most significant factor affecting measurements of weighted CTDI (CTDI_w) and a 100mm chamber would record 50-61% of the dose measured with a 450mm one. The cumulative dose measured at the centre of a 150mm long body phantom records over 70% of the equilibrium dose from a helical scan of a longer phantom. For routine CT dosimetry tests, the determination of correction factors could allow measurements with a 100mm chamber to be used to derive the CTDI that would be recorded with a longer chamber, and cumulative doses measured with a 20mm chamber in shorter phantoms to be used to calculate equilibrium doses for helical scans.
机译:全色变色膜已用于测量体模内的计算机断层扫描(CT)剂量分布。膜在来自表面治疗单元的光束中进行了校准,并通过与20mm长电离室的测量结果进行比较来确认其准确性。该结果已用于研究CT剂量测定的方法。剂量分布记录在标准CT头部和身体体模中,而散射尾部数据适合指数功能,并外推以预测较长体模中的剂量水平。数据已被用于模拟不同长度电离室的CT剂量指数(CTDI)测量和20mm室用于不同长度扫描的累积剂量测量。结果表明,笔电离室的长度是影响加权CTDI(CTDI_w)测量的最重要因素,而100mm的电离室将记录使用450mm的电离室测得的剂量的50-61%。在150mm长的人体模型中心测量的累积剂量记录了较长的人体模型的螺旋扫描所产生的平衡剂量的70%以上。对于常规的CT剂量测定测试,校正因子的确定可以允许使用100mm腔室进行测量,以得出将在较长腔室中记录的CTDI,并使用20mm腔室在较短的体模中测量的累积剂量用于计算螺旋扫描的平衡剂量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号