Thermal neutron fluence measurement in a research reactor using thermoluminescence dosimeter TLD-600

机译:使用热致发光剂量计TLD-600的研究堆中的热中子注量测量

获取原文
获取原文并翻译 | 示例
       

摘要

A thermal neutron fluence in the range between 10(11) and 10(13) n cm(-2) in the reactor core of the Tehran research reactor has been measured using TLD-600 thermoluminescence dosimeters. After a thermal treatment of 1 h at 400 degrees C followed by 20 h cooling down to room temperature of pre-exposed dosimeters in the reactor, the accumulated TL light was measured after periods of storage of 24, 48 and 72 h. The influence of the irradiation-induced damage effect on the response of TLDs and their subsequent readings has been minimized in this manner. The induced TL light due to self-activity in the TLD-600 dosimeters, which is dependent on the neutron fluence, caused a conveniently measurable TL glow curve. The induced TL in the dosimeter due to the Q-value for the beta-decay of tritium E beta-max = 18.6 keV has been reproduced separately by a beta source to check the proportions of radionuclides in the chip. A short theoretical treatment is also presented.
机译:使用TLD-600热致发光剂量计测量了德黑兰研究堆反应堆堆芯中热中子注量在10(11)和10(13)n cm(-2)之间。在400摄氏度下进行1小时的热处理,然后在反应器中冷却至室温后,将其暴露于预先暴露的剂量计的室温下20小时后,在储存24、48和72小时后,测量了累积的TL光。通过这种方式,可以将辐照引起的损伤效应对TLD响应及其后续读数的影响降到最低。由于TLD-600剂量计中的自活动性(取决于中子注量),诱导的TL光引起了可方便测量的TL发光曲线。由于source E的β衰变的Q值β-max= 18.6 keV而在剂量计中引起的TL已由β源单独复制以检查芯片中放射性核素的比例。还提供了简短的理论处理。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号