A portable system, using electrostatic collection, for the measurement of exhaled thoron activity in humans is described, together with details of the basic theory, equipment, calibration procedures, measurement and preliminary use. The portable system built on experience at the Argonne National Laboratory (ANL) to achieve a reduction in measurement time from 30 h to 200 min, and to increase the total efficiency of the system from 50% (at ANL) to 70%. The limit of detection of the system (2 sigma above zero) is 0.007 Bq of thoron activity outside the body. The total standard error of this system is 47% for a thorium lung burden of 0.22 Bq. The average background of this scintillation detector was 0.003 counts min(-1).
机译:描述了一种便携式系统,该系统使用静电收集来测量人体呼出的气活动,并详细介绍了基本理论,设备,校准程序,测量和初步使用。便携式系统建立在阿贡国家实验室(ANL)的经验基础上,可将测量时间从30小时减少到200分钟,并将系统的总效率从50%(在ANL)提高到70%。该系统的检测极限(高于零的2个sigma)为体外0.007 Bq的离子活度。对于0.22 Bq的th肺负荷,该系统的总标准误差为47%。该闪烁检测器的平均背景为0.003个计数min(-1)。