Calibration of the LNMRI secondary standard ionization chamber for ~(131)I capsules used in nuclear medicine

机译:用于核医学中〜(131)I胶囊的LNMRI二级标准电离室的校准

获取原文
获取原文并翻译 | 示例
           

摘要

The radionuclide ~(131)I has been increasingly used in nuclear medicine therapy procedures. Nowadays, the ~(131)I source administered to the patient is manufactured in two different geometries: solution and capsules. The purpose of this study is the accurate measurement of the activity present in a ~(131)I capsule without destroys it. The methodology to determine the capsules activity is to obtain the calibration factor of an IG12 secondary standard activity measurement system based on the IG12 well-type ionization chamber set up at Brazilian national metrology laboratory for ionizing radiation (LNMRI) of institute of radiation protection and dosimetry (IRD).The result obtained, 6.4670 ± 0.0381 × 10 ~(-18) A Bq ~(-1), is quite similar to the calibration factor of the ~(131)I solution contained in the standard ampoule geometry, 6.4515 ± 0.0368 × 10 ~(-18) A Bq ~(-1). After obtaining the calibration factor it was used to measure ~(131)I therapy capsules in order to check the performance of radionuclide calibrators of some Brazilian nuclear medicine centers.
机译:放射性核素〜(131)I已越来越多地用于核医学治疗程序中。如今,以两种不同的几何形状制造用于患者的〜(131)I来源:溶液和胶囊。这项研究的目的是准确测量〜(131)I胶囊中存在的活性而不会破坏它。确定胶囊活性的方法是,根据巴西辐射防护与剂量研究所的国家计量实验室(巴西国家计量实验室)建立的IG12井型电离室,获取IG12二级标准活性测量系统的校准因子。得到的结果为6.4670±0.0381×10〜(-18)A Bq〜(-1),与标准安瓿瓶几何形状中的〜(131)I溶液的校正因子6.4515±十分相似0.0368×10〜(-18)A Bq〜(-1)。获得校准因子后,它被用于测量〜(131)I治疗胶囊,以检查巴西一些核医学中心的放射性核素校准剂的性能。

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号