Comparison of posterior corneal imaging before and after LASIK using dual rotating scheimpflug and scanning slit-beam corneal tomography systems

机译:双旋转scheimpflug和扫描缝束角膜断层扫描系统对LASIK手术前后角膜后成像的比较

获取原文
获取原文并翻译 | 示例

摘要

PURPOSE: To compare the maximum posterior elevation (MPE) measurements before and after LASIK using a dual rotating Scheimpflug (DRS) imaging system (Galilei, Ziemer Ophthalmic Systems, Port, Switzerland) and a scanning slit-beam (SSB) imaging system (Orbscan IIz, Bausch & Lomb, Rochester, NY). METHODS: This retrospective study included 78 eyes from 78 patients who underwent myopic LASIK. Preoperative and postoperative data collected included anterior and posterior best-fit sphere radius and axial curvature readings, posterior central elevation (PCE), and MPE relative to a best-fit sphere using a 7.8-mm region of interest. Data were compared using paired t test analysis. RESULTS: Mean preoperative PCE (5.06 ?? 2.29 ??m with the DRS system and 12.78 ?? 6.90 ??m with the SSB system) and MPE (4.87 ?? 4 ??m with the DRS system and 15.44 ?? 9.78 ??m with the SSB system) were statistically different (P < .001). Mean postoperative PCE (4.55 ?? 2.34 ??m with the DRS system and 20.59 ?? 8.11 ??m with the SSB system) and MPE (4.90 ?? 3.35 ??m with the DRS system and 24.95 ?? 10.15 ??m with the SSB system) were statistically different (P < .001). The difference between preoperative and postoperative MPE measurements by DRS was not statistically significant (P = .953), whereas the difference measured by SSB was statistically significant (P < .001). CONCLUSIONS: The consistency of DRS measurements suggests that the posterior surface of the cornea does not change appreciably after keratorefractive surgery and is imaged more accurately using DRS compared with SSB. The DRS system affords confidence in interpreting data that are useful for discerning morphologic abnormalities of the cornea, both before and after keratorefractive surgery.
机译:目的:使用双旋转Scheimpflug(DRS)成像系统(Galilei,Ziemer Ophthalmic Systems,Port,瑞士)和扫描狭缝束(SSB)成像系统(Orbscan),比较LASIK术前和术后的最大后仰(MPE)测量值IIz,博士伦(Bausch&Lomb),纽约州罗彻斯特市)。方法:这项回顾性研究包括来自78名接受近视LASIK手术的患者的78只眼睛。收集的术前和术后数据包括前,后最佳适应球体半径和轴向曲率读数,后中央仰角(PCE)和相对于使用7.8 mm感兴趣区域的最佳适应球体的MPE。使用配对t检验分析比较数据。结果:术前平均PCE(DRS系统为5.06≤2.29 m,SSB系统为12.78≤6.90 m)和MPE(DRS系统为4.87≤4 m,15.44 9.7.7 m。 SSB系统的?m值在统计学上有差异(P <.001)。术后平均PCE(DRS系统为4.55°C 2.34 mm,SSB系统为20.59°C 8.11 mm)和MPE(DRS系统为4.90°C 3.35 mm,24.95°C 10.15 mm。 SSB系统)在统计上有差异(P <.001)。 DRS术前和术后MPE测量之间的差异无统计学意义(P = .953),而SSB所测量的差异具有统计学意义(P <.001)。结论:DRS测量的一致性表明,角膜屈光手术后角膜的后表面没有明显变化,并且与SSB相比,使用DRS可以更准确地成像。 DRS系统为解释在角膜屈光手术之前和之后对识别角膜形态异常有用的数据提供了信心。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号