首页> 外文期刊>Journal of refractive surgery >Correlation among ocular spherical aberration, corneal spherical aberration, and corneal asphericity before and after LASIK for myopic astigmatism with the SCHWIND AMARIS platform.
【24h】

Correlation among ocular spherical aberration, corneal spherical aberration, and corneal asphericity before and after LASIK for myopic astigmatism with the SCHWIND AMARIS platform.

机译:SCHWIND AMARIS平台用于近视散光的LASIK手术前后的眼球差,角膜球差和角膜非球面度之间的相关性。

获取原文
获取原文并翻译 | 示例
       

摘要

PURPOSE: To determine the spherical wave aberration of the human eye based on corneal topography. METHODS: Based on the pre- and postoperative status of 146 consecutive eyes (median patient age 36 years), the correlations between spherical aberration and asphericity and between corneal and ocular spherical aberrations were determined using simple linear regression methods. The asphericity (Q) values for which spherical aberration equals zero as well as the reference Q values for which corneal spherical aberration equals ocular spherical aberration have been determined. Patients underwent LASIK using the AMARIS excimer laser platform (SCHWIND eye-tech-solutions). All ablations were based on aspheric aberration-neutral profiles. RESULTS: Corneal and ocular spherical aberrations correlate well with Q value and the value p . R(-3) in patients before and after LASIK for myopic astigmatism. A Q value of -0.19 to -0.27 can provide zero ocular spherical aberration in patients before and after LASIK for myopic astigmatism. Ocular spherical aberration is induced at a rate of half the induced corneal spherical aberration. A reference Q value of -0.12 to +0.01 can provide corneal spherical aberration equal to ocular spherical aberration in patients before and after LASIK for myopic astigmatism. CONCLUSIONS: Ocular and corneal wave aberrations are two different concepts that are not interchangeable. As for spherical aberration, a simple static model with a reference cornea deviating from a Cartesian oval can provide a 2:1 correspondence between corneal and ocular spherical aberration.
机译:目的:基于角膜地形图确定人眼的球面波像差。方法:根据146只连续眼睛(中位年龄36岁)的术前和术后状态,使用简单的线性回归方法确定球差与非球面度以及角膜和眼球差之间的相关性。已确定球差等于零的非球面度(Q)值以及角膜球差等于眼球差的参考Q值。患者使用AMARIS准分子激光平台(SCHWIND eye-tech-solutions)进行LASIK。所有消融均基于非球面像差中性曲线。结果:角膜和眼球差与Q值和p值相关性很好。 R(-3)在LASIK手术前后的近视散光中。 Q值在-0.19至-0.27之间可为近视散光在LASIK手术前后为患者提供零眼球差。眼球差以一半的角膜球差引起。对于近视散光,LASIK手术前后患者的参考Q值-0.12至+0.01可以提供等于眼球差的角膜球差。结论:眼波和角膜波像差是两个不可互换的概念。对于球差,简单的静态模型具有参考角膜偏离笛卡尔椭圆形的特性,可以在角膜和眼球差之间提供2:1的对应关系。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号