首页> 外文期刊>Journal of refractive surgery >Incidence of complications during flap creation in LASIK using the NIDEK MK-2000 microkeratome in 26,600 cases.
【24h】

Incidence of complications during flap creation in LASIK using the NIDEK MK-2000 microkeratome in 26,600 cases.

机译:使用NIDEK MK-2000微型角膜刀在LASIK瓣制作过程中并发症的发生率为26,600例。

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

PURPOSE: To evaluate the incidence of intraoperative complications using the NIDEK MK-2000 microkeratome during flap creation in LASIK. METHODS: The incidence of intraoperative flap complications during LASIK using the NIDEK MK-2000 microkeratome was retrospectively studied in 26,600 procedures performed in 5 outpatient excimer laser surgery centers. RESULTS: A total of 65 (0.244%) complications were identified: 23 (0.086%) eyes had free caps, 13 (0.049%) eyes had an incomplete pass, 13 (0.049%) eyes had an epithelial defect, 11 (0.041%) eyes had buttonhole, and 5 (0.019%) eyes had irregular flap. The remaining 26,535 (99.756%) eyes had uneventful flap creation. CONCLUSIONS: Use of the NIDEK MK-2000 microkeratome resulted in a low incidence of intraoperative complications, making it a safe and reliable device for creating flaps during LASIK.
机译:目的:评估在LASIK皮瓣制作过程中使用NIDEK MK-2000微型角膜刀术中并发症的发生率。方法:回顾性研究了在5个门诊准分子激光手术中心进行的26,600例手术中,使用NIDEK MK-2000微型角膜刀在LASIK手术中发生皮瓣并发症的发生率。结果:总共鉴定出65例(0.244%)并发症:23例(0.086%)的患者有无顶帽手术,13例(0.049%)的患者不通过手术,13例(0.049%)的患者有上皮缺损,11例(0.041%) )眼睛有纽孔,有5眼(0.019%)眼球不规则。其余26,535眼(99.756%)的皮瓣平滑无瑕。结论:NIDEK MK-2000微型角膜刀的使用降低了术中并发症的发生率,使其成为在LASIK术中制作皮瓣的安全可靠的装置。