首页> 外文期刊>Journal of refractive surgery >Precision of higher order aberration repeatability with NIDEK OPD-scan retinoscopic aberrometry.
【24h】

Precision of higher order aberration repeatability with NIDEK OPD-scan retinoscopic aberrometry.

机译:NIDEK OPD扫描视网膜镜像差仪可实现更高阶像差重复精度。

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

PURPOSE: To evaluate repeatability of the total high order aberrations with a retinoscopic wavefront sensor. METHODS: This prospective case series analyzed 12 eyes from 6 patients who underwent wavefront measurement using retinoscopic aberrometry with the NIDEK Optical Path Difference Scan (OPD-Scan). Four consecutive wavefront measurements of each eye were taken by two trained examiners for 5.8+/-0.8-mm and 8.2+/-0.6-mm pupils (P=.002, Wilcoxon test). Total high order aberrations out to the eighth order were assessed including third order coma, third order trefoil, fourth order quadrafoil, fourth order secondary astigmatism, and fourth order spherical aberrations. Differences between measurements of all of the variables were analyzed. A P value <.05 was considered statistically significant. RESULTS: Repeatability analysis of the root-mean-square of total higher order aberrations, coma, trefoil, quadrafoil, secondary astigmatism, and spherical aberrations for both dilated and nondilated pupils did not show a statistically significant difference among all repeated measurements, except for trefoil and secondary astigmatism (analysis of variance and the Friedman test). The repeatability of total higher order aberrations was 0.15 microm for nondilated pupils and 0.18 microm for dilated pupils. Except for trefoil measurements, all high order aberrations showed reproducibility >0.15 microm when Zernike coefficients were analyzed individually. CONCLUSIONS: The NIDEK OPD-Scan aberrometer measures total higher order astigmatism and most individual aberrations with acceptable repeatability. However, measurement of trefoil with this instrument is less repeatable.
机译:目的:使用视网膜镜波前传感器评估总高阶像差的可重复性。方法:该前瞻性病例系列分析了6例患者的12眼,这些患者使用NIDEK光程差扫描(OPD-Scan)使用视网膜镜像差法进行波前测量。两名训练有素的检查者对每只眼睛的5.8 +/- 0.8mm和8.2 +/- 0.6mm瞳孔进行了连续四次波前测量(P = .002,Wilcoxon测试)。评估了到八阶的总高阶像差,包括三阶彗差,三阶三叶形,四阶四翼形,四阶二次像散和四阶球差。分析所有变量的测量值之间的差异。 P值<.05被认为具有统计学意义。结果:对散瞳和非散瞳的总高阶像差,彗差,三叶形,四翼形,二次散光和球面像差的均方根进行重复性分析,除三叶形外,所有重复测量均未显示出统计学上的显着差异和继发性散光(方差分析和Friedman检验)。总高阶像差的可重复性对于非散瞳为0.15微米,对散瞳为0.18微米。除三叶测量外,当分别分析Zernike系数时,所有高阶像差均显示可重复性> 0.15微米。结论:NIDEK OPD-Scan像差仪可测量总体高阶像散和大多数个体像差,并且具有可重复性。但是,用此仪器测量三叶草的重复性较差。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号