首页> 外文期刊>Journal of Raman Spectroscopy: An International Journal for Original Work in All Aspects of Raman Spectroscopy, Including Higher Order Processes, and Also Brillouin- and Rayleigh Scattering >Quick detection of contaminants leaching from polypropylene centrifuge tubes with surface-enhanced Raman spectroscopy and ultraviolet absorption spectroscopy
【24h】

Quick detection of contaminants leaching from polypropylene centrifuge tubes with surface-enhanced Raman spectroscopy and ultraviolet absorption spectroscopy

机译:表面增强拉曼光谱和紫外吸收光谱快速检测从聚丙烯离心管中滤出的污染物

获取原文
获取原文并翻译 | 示例
           

摘要

Anomalous surface-enhanced Raman scattering (SERS) peaks were identified for liquid sample stored in polypropylene (PP) centrifuge tubes for months. We observed unexpected Raman peaks during experiments with thiamine hydrochloride aqueous solutions stored in PP tubes for 2 months. In order to identify the contaminants, we have performed SERS experiments on deionized (DI) water stored in PP centrifuge tubes for 2 months and compared them with those from fresh DI water sample. We have also carried out ultraviolet (UV) absorption spectra for both fresh and contaminated water. We believe that the water is contaminated because of chemicals leaching from the PP tube. From the gas chromatography-mass spectrometry data, the main contaminants were found to be phthalic acid (PA) and its derivatives. Further SERS and UV absorption experiment for PA correlated well with the anomalous peaks identified earlier. We qualitatively confirmed the identification and quantitatively estimated the concentration of the suspect contaminants as between 1 and 10 μM with both SERS and UV absorption spectroscopy. With UV absorption spectroscopy, we precisely estimated the concentration as 2.1 μM. We have shown that the sample in PP tube can be contaminated by the leaching chemicals upon long-term storage, and suggest SERS and UV absorption spectroscopy as two quick and simple techniques to detect the contamination.
机译:对于在聚丙烯(PP)离心管中存储了数月的液体样品,发现了异常的表面增强拉曼散射(SERS)峰。在将硫胺素盐酸盐水溶液储存在PP管中2个月的实验期间,我们观察到了意外的拉曼峰。为了识别污染物,我们对存储在PP离心管中的去离子(DI)水进行了2个月的SERS实验,并将其与新鲜DI水样品中的去离子水进行了比较。我们还对淡水和污水进行了紫外线(UV)吸收光谱分析。我们认为,由于PP管中的化学物质浸出,水受到了污染。根据气相色谱-质谱数据,发现主要污染物为邻苯二甲酸(PA)及其衍生物。进一步的PA的SERS和UV吸收实验与较早发现的异常峰有很好的相关性。我们通过SERS和UV吸收光谱学定性地确认了鉴定并定量估计了1-10μM的可疑污染物浓度。使用紫外吸收光谱法,我们精确地估计浓度为2.1μM。我们已经表明,PP管中的样品在长期存储时会被浸出的化学物质污染,因此建议将SERS和UV吸收光谱法作为检测污染物的两种快速而简单的技术。

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号