Single versus double ended high frequency pressurization of mock arteries: symmetry of expansion.

机译:单端和双端高频模拟动脉加压:扩张的对称性。

获取原文
获取原文并翻译 | 示例
       

摘要

A series of experiments was carried out to evaluate the expansion characteristics of silicone mock arteries at frequencies from 5 to 100 Hz. Of particular interest was the magnitude of expansion along the length of the tube versus frequency. The tubes were pressurized from one end or from both ends. In general, results indicate asymmetric expansion along the length of the tube. This asymmetry, however, changed dramatically with frequency. These changes are not fully explainable by standard experimental considerations such as compliance and harmonics.
机译:进行了一系列实验,以评估硅酮模拟动脉在5至100 Hz频率下的扩张特性。特别令人感兴趣的是沿着管的长度的膨胀量与频率的关系。从一端或从两端对管加压。通常,结果表明沿管的长度不对称膨胀。但是,这种不对称性随频率发生了巨大变化。这些变化无法通过标准实验考虑因素(如柔度和谐波)完全解释。