Effectiveness of a breath-hold monitoring system in improving the reproducibility of different breath-hold positions in multiphasic CT imaging

机译:屏气监测系统在多相CT成像中提高不同屏气位置可重复性的有效性

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

This study tests whether the utilization of an electronic breath-hold monitoring device improves breath-hold reproducibility during computed tomographic (CT) scanning. Two cohorts of 40 patients underwent dual-phase abdominal CT scans, either with a breath-hold monitoring device or with the standard breath-holding technique. Two blinded readers measured the differences in diaphragmatic position between phases. There was no statistical difference in diaphragmatic position (P=.14) between the monitored (8.5±11.5 mm) and control (5.6±5.2 mm) cohorts. Ten percent of patients from the monitored cohort had greater than 20 mm of deviation, versus 0%-2.5% for the control cohort. Reproduction of breath-holding position remains challenging, even with a monitoring system.
机译:这项研究测试了在计算机断层扫描(CT)扫描过程中使用电子屏气监测设备是否可以提高屏气再现性。两个队列的40名患者接受了屏气监测设备或标准屏气技术进行了双阶段腹部CT扫描。两个盲的阅读器测量了相之间隔膜位置的差异。在监测组(8.5±11.5 mm)和对照组(5.6±5.2 mm)之间,diaphragm肌位置(P = .14)无统计学差异。监测队列中有10%的患者偏差大于20 mm,而对照队列中则为0%-2.5%。即使使用监控系统,屏住呼吸的姿势仍然具有挑战性。