Testing a dual-Modality system that combines full-field digital mammography and automated breast ultrasound

机译:测试将全视野数字化乳腺X射线摄影术和自动乳房超声检查相结合的双模系统

获取原文
获取原文并翻译 | 示例
       

摘要

Purpose: The aim of this study was to test a novel dual-modality imaging system that combines full-field digital mammography (FFDM) and automated breast ultrasound (ABUS) in a single platform. Our Aceso system, named after the Greek goddess of healing, was specifically designed for the early detection of cancer in women with dense breast tissue. Materials and Methods: Aceso was first tested using two industry standards: a Contrast Detail Mammography (CDMAM) phantom as endorsed by European Reference Organisation for Quality Assured Breast Screening and Diagnostic Services was used to assess the FFDM images; and the CIRS 040GSE ultrasound phantom was imaged to evaluate the quality of the ABUS images. In addition, 58 women participated in a clinical trial: 51 were healthy volunteers aged between 40 and 65, while 7 were patients referred by the breast clinic, 6 of whom had biopsy-proven breast cancer. Results: The CDMAM tests showed that the FFDM results were "acceptable" but fell short of "achievable" which was attributed to the low dose used. The ABUS images had good depth penetration (80 mm) and adequate axial resolution (0.5 mm), but the lateral resolution of 2 mm was judged to be too coarse. In a 42-year-old volunteer with extremely dense breast tissue, the ABUS modality detected a lesion (a benign cyst) that was mammographically occult in the FFDM image. For a 73-year-old patient with fatty breasts, a malignant lesion was successfully detected and co-registered in the FFDM and ABUS images. On average, each woman spent less than 11 min in the acquisition room. Conclusions: While there is room for improvement in the quality of both the FFDM and ABUS images, Aceso has demonstrated its ability to acquire clinically meaningful images for a range of women with varying breast densities and, therefore, has potential as a screening device. (C) 2016 Elsevier Inc. All rights reserved.
机译:目的:这项研究的目的是测试一个新颖的双模态成像系统,该系统在一个平台上结合了全场数字化乳房X线照相术(FFDM)和自动乳房超声检查(ABUS)。我们的Aceso系统以希腊治愈女神的名字命名,专门用于早期发现乳腺组织密集的女性中的癌症。材料和方法:Aceso首先使用两种行业标准进行了测试:使用经欧洲质量保证乳房筛查和诊断服务参考组织认可的对比细节乳腺摄影(CDMAM)体模来评估FFDM图像;然后对CIRS 040GSE超声体模成像,以评估ABUS图像的质量。此外,有58名妇女参加了临床试验:51名年龄在40至65岁之间的健康志愿者,而7名是由乳房诊所转诊的患者,其中6名经过活检证实为乳腺癌。结果:CDMAM测试表明,FFDM结果“可以接受”,但由于使用的剂量低而未达到“可实现”。 ABUS图像具有良好的深度穿透(80毫米)和足够的轴向分辨率(0.5毫米),但是2毫米的横向分辨率被认为太粗糙。在一位42岁的志愿者中,乳房组织非常致密,ABUS模态检测到的病变(良性囊肿)在FFDM图像中被乳腺X线摄影所掩盖。对于一名73岁的肥大乳房患者,成功检测出恶性病变并将其共记录在FFDM和ABUS图像中。平均而言,每位女性在收养室花费的时间少于11分钟。结论:虽然FFDM和ABUS图像的质量都有改善的空间,但Aceso展示了其能够为一系列乳房密度不同的妇女获取具有临床意义的图像的能力,因此,它具有作为筛查设备的潜力。 (C)2016 Elsevier Inc.保留所有权利。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号