首页> 外文期刊>Journal of Physics, D. Applied Physics: A Europhysics Journal >HPGe detector absolute full-energy peak efficiency calibration including coincidence correction for circular disc sources
【24h】

HPGe detector absolute full-energy peak efficiency calibration including coincidence correction for circular disc sources

机译:HPGe检测器绝对全能量峰值效率校准,包括对圆盘源的重合校正

获取原文
获取原文并翻译 | 示例
           

摘要

A direct mathematical approach is presented to calculate the absolute full-energy peak (photopeak) efficiency and coincidence correction of hyper pure germanium (HPGe) detectors for extended circular disc sources. The calculation of the full-energy peak efficiency is based on elliptic integration and its validity was checked by comparison with experimental measurements. Extensive experimental readings have been carried out; gamma sources containing several radionuclides covering the energy range from 60 to 1836 keV were used. By comparison, the theoretical and experimental full-energy peak efficiency values are in good agreement; the overall percentage error is less than 3%.
机译:提出了一种直接的数学方法来计算用于扩展圆盘源的绝对全能峰(光峰)效率和超纯锗(HPGe)探测器的重合校正。全能量峰值效率的计算基于椭圆积分,并通过与实验测量值的比较来检验其有效性。已经进行了广泛的实验阅读;使用了包含数种放射性核素的伽马射线源,其能量范围从60到1836 keV。相比之下,理论和实验全能量峰值效率值吻合良好;总体百分比误差小于3%。