Three-dimensional imaging of the spine using the EOS system: Is it reliable? A comparative study using computed tomography imaging

机译:使用EOS系统对脊柱进行三维成像:是否可靠?使用计算机断层扫描成像的比较研究

获取原文
获取原文并翻译 | 示例
           

摘要

The aim of this study was to evaluate the precision of three-dimensional geometry compared with computed tomography (CT) images. This retrospective study included patients who had undergone both imaging of the spine using the EOS imaging system and CT scanning of the spine. The apical vertebral orientation was also measured using the EOS imaging system and by CT. Other measures such as the Cobb angle and apical vertebral rotation and translation were used as the control variables to evaluate the potential discrepancy between the standing position in EOS imaging and the supine position in CT scanning. The apical vertebral orientations were 8.7 for the first measurement and 8.4 for the second measurement made by the first author, and 10.3 for the measurement made by the second author. The average of these measurements was 9.3 compared with 6.6 (P=0.65) obtained on CT scanning. The precision of EOS-based measurements of vertebral rotation has never been tested in clinical practice. Although it has limitations, this study suggests that the results obtained using EOS are comparable to those obtained on CT.
机译:这项研究的目的是评估与计算机断层扫描(CT)图像相比的三维几何形状的精度。这项回顾性研究包括既使用EOS成像系统进行脊柱成像又进行CT扫描的患者。还使用EOS成像系统和CT测量了椎骨的顶部方向。使用其他测量方法(例如Cobb角和椎骨顶尖旋转和平移)作为控制变量,以评估EOS成像中站立位置与CT扫描中仰卧位置之间的潜在差异。第一作者的第二次测量的根尖椎骨方向为8.7,第二次测量为8.4,第二作者的测量根尖椎骨的方向为10.3。这些测量的平均值为9.3,而CT扫描的平均值为6.6(P = 0.65)。从未在临床实践中测试过基于EOS的椎骨旋转测量的精度。尽管它有局限性,但这项研究表明,使用EOS所获得的结果与CT所获得的结果相当。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号