首页> 外文期刊>Journal of Pharmaceutical and Biomedical Analysis: An International Journal on All Drug-Related Topics in Pharmaceutical, Biomedical and Clinical Analysis >Differential pulse adsorption voltammetry for determination of procaine hydrochloride at a pumice modified carbon paste electrode in pharmaceutical preparations and urine.
【24h】

Differential pulse adsorption voltammetry for determination of procaine hydrochloride at a pumice modified carbon paste electrode in pharmaceutical preparations and urine.

机译:差分脉冲吸附伏安法测定药物制剂和尿液中的浮石修饰的碳糊电极上的盐酸普鲁卡因。

获取原文
获取原文并翻译 | 示例
           

摘要

Procaine hydrochloride was determined by the differential pulse voltammetry (DPV) using a 6% (m/m) pumice modified carbon paste electrode in 1.25x10(-3) mol l(-1) KH(2)PO(4) and Na(2)HPO(4) buffer solution (pH 6.88, 25 degrees C). The anodic peak potential used was +0.980 V (vs. SCE). A good linear relationship was realized between the anodic peak current and procaine concentration in the range of 9.0x10(-7)-2.6x10(-5) mol l(-1) with the detection limit of 5.0x10(-8) mol l(-1). The recovery was 95.2-104.8% with the relative standard deviation of 3.2% (n=10). The pharmaceutical preparations, procaine hydrochloride injection and the urine samples were determined with the desirable results.