Pullout strengths of self-reinforced poly-L-lactide (SR-PLLA) rods versus Kirschner wires in bovine femur.

机译:牛股骨中自增强聚L-丙交酯(SR-PLLA)棒与Kirschner线的抗拉强度。

获取原文
获取原文并翻译 | 示例
           

摘要

OBJECTIVE: To determine the relative amount of fixation of self-reinforced poly-L-lactide (SR-PLLA) rods and Kirschner wires in bovine cancellous bone by comparing their pullout strength DESIGN: An in vitro laboratory study was performed using bovine femurs. Ten two-millimeter-diameter pins of each type were inserted into cancellous bone and then pulled out, using a material testing machine. The maximum force (pullout strength) was selected over other measurements to compare the amount of fixation of the two types of pins. All of the pins were retrieved for microscopic analysis. A paired t test was performed to analyze the differences between the pullout strength of the two types of pins. SETTING: Orthopaedic Bioengineering Laboratory, University of Louisville School of Medicine, Louisville, Kentucky, U.S.A. SPECIMENS: Two young fresh bovine distal femurs, ten two-millimeter-diameter Kirschner wires, ten two-millimeter-diameter bioabsorbable SR-PLLA rods MAIN OUTCOME MEASUREMENTS: Pullout strength in Newtons, and microscopic pin surface aspect after insertion. RESULTS: Significant differences were noted between the maximum force required to remove the two types of pins (p < 0.01) The K-wire mean pullout force was 37.7 N (SD 13.6), and the SR-PLLA rod mean pullout force was 53.6 N (SD 19.3). Microscopic analysis indicated surface modification only on the SR PLLA rods. DISCUSSION: SR-PLLA composites have shown comparable clinical results to their metallic counterparts. In this study, the pullout strength of SR-PLLA rods was compared with that of conventional K-wires. A significant difference (p < 0.01) favoring bioabsorbable pullout strength was noted. The bioabsorbable pin surface modification during insertion is an interesting finding that warrants further investigation as a potential source of improved fixation. CONCLUSION: SR-PLLA rods retain their hold in bovine cancellous bone better than K-wires. This finding offers to the orthopaedic surgeon more information about new pin fixation methods.
机译:目的:通过比较牛股骨松质骨的拔出强度来确定自增强聚L-丙交酯(SR-PLLA)棒和克氏针的固定量。设计:使用牛股骨进行了体外实验室研究。将每种类型的十个两毫米直径的销钉插入松质骨中,然后使用材料测试机将其拔出。选择最大的力(拉出强度)来进行其他测量,以比较两种类型的销钉的固定量。回收所有销以进行显微分析。进行配对t检验以分析两种类型的销的拉拔强度之间的差异。地点:路易斯维尔大学医学院骨科生物工程实验室,美国肯塔基州路易斯维尔规格:两只新鲜的新鲜牛股骨远端,十根直径为两毫米的克氏针,十根直径为两毫米的可生物吸收的SR-PLLA棒主要结果:插入后的拔出强度(以牛顿计)和细微的针表面。结果:移除两种类型的销所需的最大力之间存在显着差异(p <0.01)。K线平均拔出力为37.7 N(SD 13.6),而SR-PLLA杆平均拔出力为53.6 N (SD 19.3)。显微分析表明仅在SR PLLA棒上进行了表面改性。讨论:SR-PLLA复合材料已显示出与金属同类产品相当的临床结果。在这项研究中,将SR-PLLA棒的拔出强度与常规K线的拔出强度进行了比较。注意到有利于生物吸收的拉出强度的显着差异(p <0.01)。插入过程中可生物吸收的针表面改性是一个有趣的发现,值得进一步研究,以作为改进固定的潜在来源。结论:SR-PLLA棒比K-wire更好地将其保持在牛松质骨中。这一发现为整形外科医生提供了有关新的销钉固定方法的更多信息。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号