MRI study of bioabsorbable poly-L-lactic acid devices used for fixation of fracture and osteotomies.

机译:用于骨折和截骨固定的生物可吸收聚L-乳酸装置的MRI研究。

获取原文
获取原文并翻译 | 示例
           

摘要

BACKGROUND: The overall clinical results of bioabsorbable fixation devices made of poly-L-lactic acid (PLLA) used for fixation of fractures, bone grafting, and osteotomies have been favorable. However, clinical studies demonstrated no sign of normal bony architecture restored after surgery, although implant channels had been filled with fibrous tissue. The purpose of the present retrospective study was to examine the extent of structural changes in PLLA devices (PLLA-Ds) for fixation of rotational acetabular osteotomies and displaced malleolar ankle fractures using magnetic resonance imaging (MRI). METHODS: Altogether, 14 patients with osteoarthritis of hip joints and 15 with displaced malleolar ankle fractures were operated on using PLLA-D (NEOFIX). Of these patients, 22 were finally enrolled in the study, and the period from operation to the time of the study ranged from 17 to 78 months. The postoperative radiographic findings were evaluated for union, and changes around the implant holes were classified as sclerosis, resorption, or no change. MRI was carried out to estimate changes in the PLLA-Ds. RESULTS: Bone union was obtained in all cases; clinical complications such as infection, joint effusion, soft tissue irritation due to PLLA-D deviation, and motion pain in the joints were not observed. The MRI study suggested that water content in PLLA-D increased mainly due to biodegradation and that implants were not replaced by bony tissue. CONCLUSIONS: The PLLA-Ds were degraded but were not replaced by bony tissue during the observation period. Considering these findings and the assumption that in bony tissues mechanical strength of PLLA-D decreases with time, attention should be paid to mechanical insufficiency, which may occur when the cross-sectional area of a PLLA-D extends beyond the cross-sectional area of the osteosynthesis site.
机译:背景:用于骨折,植骨和截骨术的聚L-乳酸(PLLA)制成的生物可吸收固定装置的整体临床效果良好。但是,临床研究表明,尽管植入物通道已充满纤维组织,但手术后没有恢复正常骨结构的迹象。本回顾性研究的目的是使用磁共振成像(MRI)检查固定旋转髋臼截骨术和移位的踝踝骨折的PLLA设备(PLLA-Ds)的结构变化程度。方法:共使用PLLA-D(NEOFIX)手术治疗14例髋关节骨关节炎和15例踝踝移位的骨折。在这些患者中,最终有22名患者参加了研究,从手术到研究时间为17到78个月不等。评估术后影像学检查的结合,并将植入孔周围的变化分类为硬化,吸收或无变化。进行MRI以估计PLLA-D的变化。结果:所有病例均获得了骨结合。未观察到临床并发症,如感染,关节积液,由于PLLA-D偏差引起的软组织刺激和关节运动疼痛。 MRI研究表明,PLLA-D中的水含量增加主要是由于生物降解,并且植入物没有被骨组织替代。结论:在观察期间,PLLA-Ds被降解但未被骨组织替代。考虑到这些发现,并假设在骨组织中PLLA-D的机械强度会随着时间的推移而降低,应注意机械功能不全,当PLLA-D的横截面面积超出其横截面时,可能会发生机械性不足。骨合成部位。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号