A new infusion pathway monitoring system utilizing electrostatic induced potential.

机译:一种利用静电感应电势的新型输液路径监测系统。

获取原文
获取原文并翻译 | 示例
           

摘要

We have developed a new infusion pathway monitoring system employing linear integrated circuits and a low-power 8-bit single chip microcomputer. The system is available for hospital and home use and it constantly monitors the intactness of the pathway. The sensor is an electro-conductive polymer electrode wrapped around the infusion polyvinyl chloride infusion tube. This records an AC (alternating current) voltage induced on the patient's body by electrostatic coupling from the normal 100 volt, 60 Hz AC power line wiring field in the patient's room. If the injection needle or infusion tube becomes detached, then the system detects changes in the induced AC voltage and alerts the nursing station, via the nurse call system or PHS (personal handy phone System).
机译:我们已经开发出了一种新的采用线性集成电路和低功耗8位单片机的输液通道监控系统。该系统可用于医院和家庭,并不断监控通道的完整性。传感器是包裹在输液聚氯乙烯输液管周围的导电聚合物电极。它记录了通过静电耦合从患者房间的正常100伏,60赫兹交流电源线的布线场在患者身体上感应出的交流(交流)电压。如果注射针或输液管松开,则系统会检测出感应交流电压的变化,并通过护士呼叫系统或PHS(个人手持电话系统)向护理站发出警报。