X-ray microanalysis of non-biological and biological specimens was carried out in the environmental scanning electron microscope (ESEM) under different conditions of specimen distance (the distance travelled by the electron probe within the specimen chamber) and chamber atmosphere. Using both water vapour and argon atmospheres, it was shown that reduction in specimen distance had no effect on atmospheric gas X-ray signal in either case. Unlike water vapour, increased levels of argon (up to 10 torr) caused a marked depression of specimen P/B ratios, with a decrease in both characteristic and background (continuum) counts, These effects in argon were not altered by reduction in specimen distance. Specimen distance was important in relation to beam skirting and elemental analysis. With an extended assembly (short specimen distance), beam skirting in a water-vapour atmosphere was much reduced - leading to enhanced element detectability in a discrete biological specimen (Anabaena cyclindrica). [References: 21]
机译:非生物和生物样品的X射线显微分析是在环境扫描电子显微镜(ESEM)中在不同的样品距离(电子探针在样品腔内移动的距离)和腔室气氛的不同条件下进行的。使用水蒸气和氩气气氛,都表明减小样品距离对两种情况下的大气X射线信号都没有影响。与水蒸气不同,增加的氩气含量(最多10托)导致样品P / B比明显降低,特征和背景(连续)数均降低。这些对氩气的影响不会因样品距离的减小而改变。标本距离对于束裙部和元素分析很重要。通过扩展的组件(较短的标本距离),水蒸气气氛中的束流裙边大大减少了-从而提高了离散生物标本(Anabaena cyclindrica)中元素的可检测性。 [参考:21]

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号