Potential for on-site analysis of hog manure using a visual and near infrared diode array reflectance spectrometer

机译:使用可见和近红外二极管阵列反射光谱仪对猪粪进行现场分析的潜力

获取原文
获取原文并翻译 | 示例

摘要

In this study, a diode array near infrared (MR) spectrometer that can be mounted on a mobile agricultural machine was tested under lab conditions for its effectiveness as a rapid, on-site hog (Sus domesticus) manure composition sensor. One hundred and sixty nine diverse hog manures, collected from farms in the northern part of Belgium, were assayed by conventional wet chemical analysis and spectroscopy for the following constituents: dry matter content, organic matter content, pH, total Kjeldahl nitrogen, ammonium nitrogen, phosphorus, potash, calcium, sodium and magnesium. Samples were scanned (two fills, three measurements each) on a Zeiss Corona 1.7 VisNIR diode array instrument equipped with an OMK measuring head, ranging between 400 and 1700 nm (five scans per measurement) using a sample cell (50mm diameter) for diffuse reflectance scanning from the bottom. Based on the absorbance-constituent and constituent-constituent correlations, the constituents were divided into two groups: one for those in the solids (dung) and one for those in the liquid (urine). The pH was not correlated to any other constituent and, therefore, treated separately. Partial least squares calibrations were developed for the constituent groups and for each of the constituents separately using leave-one-out cross-validation, showing a common spectral basis for each constituent group. To investigate the possibility of developing a cost-effective sensor, regression calibrations were developed using a few wavelengths selected based on the partial least squares regression and validated by leave-one-out cross-validation. Results demonstrated that: (1) approximate quantitative predictions can be made for total N, NH{sub}4-N and K, (2) a discrimination between high and low values can be made for DM, OM, P and Mg, (3) prediction of Ca, Na and pH are not possible (yet), (4) one measurement and one filling of the cup give a reliable measurement, (5) a cost-effective sensor measuring only a few appropriately selected wavelengths might be able to give comparable results, (6) further work will be necessary to optimise sample presentation and calibration procedures and (7) mounting a diode array instrument on a slurry tanker can make environmentally and economically sound nutrient management possible.
机译:在这项研究中,可以在实验室条件下测试可以安装在移动农业机械上的二极管阵列近红外(MR)光谱仪作为快速现场猪(Sus domesticus)粪便成分传感器的有效性。通过常规湿化学分析和光谱法分析了从比利时北部农场收集的169种猪肥中的以下成分:干物质含量,有机物质含量,pH,凯氏定氮,氨氮,磷,钾,钙,钠和镁。使用配备了OMK测量头的Zeiss Corona 1.7 VisNIR二极管阵列仪器对样品进行扫描(两次填充,每次三个测量),使用样品池(直径50mm)进行漫反射,扫描范围为400至1700 nm(每次测量五次扫描)从底部扫描。根据吸光度成分和成分成分的相关性,将成分分为两组:一组用于固体(粪便),另一种用于液体(尿液)。 pH与其他任何成分都不相关,因此需要单独处理。使用留一法交叉验证为构成组和每种构成分别开发了偏最小二乘校正,显示了每个构成组的通用光谱基础。为了研究开发具有成本效益的传感器的可能性,使用了基于偏最小二乘回归选择的几个波长来进行回归校准,并通过留一法交叉验证进行了验证。结果表明:(1)可以对总N,NH {sub} 4-N和K进行近似的定量预测,(2)可以对DM,OM,P和Mg进行高低值的区分,( 3)尚无法预测Ca,Na和pH(4),(4)一次测量和一次填充量杯即可提供可靠的测量结果,(5)具有成本效益的传感器,只能测量几个适当选择的波长为了获得可比的结果,(6)将需要进一步的工作来优化样品的呈现和校准程序;(7)在浆液罐车上安装二极管阵列仪器可以使对环境和经济的养分管理成为可能。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号