首页> 外文期刊>Journal of near infrared spectroscopy >Solid dosage form analysis by near infrared spectroscopy: comparison of reflectance and transmittance measurements including the determination of effective sample mass
【24h】

Solid dosage form analysis by near infrared spectroscopy: comparison of reflectance and transmittance measurements including the determination of effective sample mass

机译:通过近红外光谱对固体剂型进行分析:比较反射率和透射率,包括确定有效样品质量

获取原文
获取原文并翻译 | 示例
           

摘要

Investigations into the near infrared analysis of intact tablets allow comparison of the reflectance and transmittance methodologies. The studies involved estimation of the effective sample mass and the effect of sample inhomogeneities. An empirical method for determining effective sample mass used tablets of varying thickness and constant compression force. Tablets containing a range of cimetidine concentrations were used to study the effect of drug concentration on effective sample mass as a function of wavelength of observation. Effective sampling depth was between 1.9 and 2.7 mm for reflectance measurements. With transmittance measurements, signal was completely attenuated when tablet thickness was greater than 3.4-4.9 mm, depending on the wavelength of observation. Multi-layered tablets provided a means of probing the effect of sample inhomogeneity. Both reflectance and transmittance measurements may be sensitive to sample inhomogeneity. Transmittance measurements are sensitive to pathlength variations.
机译:对完整片剂的近红外分析的研究可以比较反射率和透射率方法。研究涉及有效样品质量的估计和样品不均匀性的影响。一种确定有效样品质量的经验方法是使用厚度和压力恒定的片剂。使用含有一系列西咪替丁浓度的片剂来研究药物浓度对有效样品质量的影响,该作用是观察波长的函数。对于反射率测量,有效采样深度在1.9到2.7毫米之间。通过透射率测量,当片剂厚度大于3.4-4.9 mm时,信号将完全衰减,具体取决于观察波长。多层片剂提供了一种探测样品不均匀性影响的方法。反射率和透射率测量都可能对样品的不均匀性敏感。透射率测量对光程变化敏感。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号