首页> 外文期刊>Journal of near infrared spectroscopy >True accuracy of near infrared spectroscopy and its dependence on precision of reference data
【24h】

True accuracy of near infrared spectroscopy and its dependence on precision of reference data

机译:近红外光谱的真实准确性及其对参考数据准确性的依赖性

获取原文
获取原文并翻译 | 示例
       

摘要

A more precise estimate of the accuracy of near infrared (NIR) spectroscopy is obtained when the measured standard errors of cross validation (SECV) and prediction (SEP) are corrected for imprecision of the reference data. The significance of correction increases with increasing imprecision of reference data. Very high precision of reference data obtained through replicate analyses under reproducibility conditions may not be the optimal goal for the development of calibration equations. In a situation of limited resources, the precision of the reference data should be related to the obtainable accuracy of the spectroscopic system. Investigation of several routine applications based on the partial least-squares (PLS) regression technique showed that increased precision of calibration data only resulted in marginal improvements in true accuracy if the total standard error of reference results from the beginning was less than the estimated true accuracy of the corresponding NIR calibration.
机译:对参考数据的不精确性进行校正的交叉验证(SECV)和预测(SEP)的标准误差校正后,可以获得对近红外(NIR)光谱准确性的更精确估计。校正的重要性随着参考数据的不精确性的增加而增加。在可重复性条件下通过重复分析获得的参考数据的非常高精度可能不是开发校准方程式的最佳目标。在资源有限的情况下,参考数据的精度应与光谱系统的可获得精度有关。对基于偏最小二乘(PLS)回归技术的几种常规应用的研究表明,如果从一开始就得出的参考结果的总标准误差小于估计的真实准确度,则增加校准数据的准确性只会导致真实准确度的边际改进。相应的NIR校准。