Osseointegration of zirconia and titanium dental implants: a histological and histomorphometrical study in the maxilla of pigs.

机译:氧化锆和钛牙科植入物的骨整合:猪上颌骨的组织学和组织形态计量学研究。

获取原文
获取原文并翻译 | 示例

摘要

OBJECTIVES: The purpose of the present study was to histologically compare the bone tissue responses to surface-modified zirconia and titanium implants. METHODS: Threaded zirconia implants were produced using a new low-pressure injection moulding technique and thereafter surface treated by acid etching. Titanium implants with the exact shape and surface treated by sandblasting and acid etching (SLA) served as controls. Fifteen adult pigs received both implant types in the maxilla 6 months after extraction of the second and third incisors. The animals were sacrificed after 4, 8 and 12 weeks and 30 implants with surrounding bone were retrieved. RESULTS: Histological evaluation showed osseous integration for both materials. Zirconia implants revealed mean peri-implant bone density values of 42.3% (SD +/- 14.5) at 4 weeks, 52.6% (SD +/- 5.7) at 8 weeks and 54.6% (SD +/- 11.5) at 12 weeks after implantation, whereas Ti-SLA implants demonstrated mean values of 29% (SD +/- 10), 44.1% (SD +/- 18) and 51.6% (SD +/- 8.6) at corresponding time intervals. With respect to the bone-implant contact ratio, the mean values for zirconia ranged between 27.1% (SD +/- 3.5) and 51.1% (SD +/- 12.4) and for Ti-SLA, it ranged between 23.5% (SD +/- 7.5) and 58.5% (SD +/- 11.4). For the parameters investigated, no statistically significant differences between both types of implants could be detected at any time point. CONCLUSIONS: No statistical difference between implants could be demonstrated with any of the methods used. The limited number of animals per group, however, does not allow to conclude that there is no difference in osseointegration between the two types of implants, although the data tend to suggest such a trend.
机译:目的:本研究的目的是从组织学上比较骨组织对表面改性的氧化锆和钛植入物的反应。方法:使用新型低压注射成型技术生产螺纹氧化锆植入物,然后通过酸蚀刻进行表面处理。通过喷砂和酸蚀(SLA)处理的具有精确形状和表面的钛植入物作为对照。在拔出第二和第三颗门齿后的六个月,有十五只成年猪在上颌接受了两种植入物类型。在第4、8和12周后处死动物,并取回30个带有周围骨骼的植入物。结果:组织学评价显示两种材料骨整合。氧化锆植入物在第4周时的平均植入物周围骨密度值为42.3%(SD +/- 14.5),在第8周时为52.6%(SD +/- 5.7),在第12周后为54.6%(SD +/- 11.5)植入,而Ti-SLA植入物在相应的时间间隔显示平均值分别为29%(SD +/- 10),44.1%(SD +/- 18)和51.6%(SD +/- 8.6)。关于骨与植入物的接触率,氧化锆的平均值在27.1%(SD +/- 3.5)和51.1%(SD +/- 12.4)之间,而Ti-SLA的平均值在23.5%(SD + /-7.5)和58.5%(SD +/- 11.4)。对于所研究的参数,在任何时间点都无法检测到两种类型的植入物之间的统计学显着差异。结论:使用任何一种方法均无法证明植入物之间的统计学差异。但是,每组动物数量有限,不能得出结论,两种类型的植入物之间的骨整合没有差异,尽管数据倾向于显示出这种趋势。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号