首页> 外文期刊>Journal of Modern Optics >Optical quality of intraocular lenses after several years in the human eye
【24h】

Optical quality of intraocular lenses after several years in the human eye

机译:数年后在人眼中人工晶状体的光学质量

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

The optical quality of explanted intraocular lenses (IOLs) was analysed to verify if their quality is maintained after having been in the human eye for several years. Twenty-five explanted IOLs were studied, 10 because of intraocular luxation and 15 from corneal donors. The dioptric power and modulation transfer function (MTF) were determined to discover the image quality of the IOLs. The spectral transmission was also established to ascertain whether the filter against ultraviolet (UV) radiation was still effective. The results show that the MTF of implanted IOLs is maintained and therefore so is the image quality they provide. Only decay in the MTF of some of the explanted IOLs was observed because of intraocular luxation. The transmission of the UV filters incorporated in the different IOLs hardly varied; thus they maintained their effectiveness.
机译:对植入式人工晶状体(IOL)的光学质量进行了分析,以验证它们在人眼中呆了几年后是否能保持其质量。研究了二十五个外植人工晶体,其中十个是由于眼内脱位,而十五个是来自角膜供体。确定屈光度和调制传递函数(MTF)以发现IOL的图像质量。还确定了光谱透射率,以确定抗紫外线(UV)辐射的滤光片是否仍然有效。结果表明,植入的IOL的MTF保持不变,因此它们提供的图像质量也保持不变。由于眼内半脱位,仅观察到某些外植IOL的MTF下降。包含在不同IOL中的UV滤光片的透射率几乎不变。因此,他们保持了效力。