Reproducibility of total cerebral blood flow measurements using phase contrast magnetic resonance imaging.

机译:使用相衬磁共振成像的全脑血流测量的可重复性。

获取原文
获取原文并翻译 | 示例
           

摘要

PurposeTo evaluate reproducibility of total cerebral blood flow (CBF) measurements with phase contrast magnetic resonance imaging (pcMRI).Materials and MethodsWe repeated total CBF measurements in 15 healthy volunteers with and without cardiac triggering, and with and without repositioning. In eight volunteers measurements were performed at two different occasions. In addition, measurement of flow in a phantom was performed to validate MR measurements.ResultsA difference of 40.4 ml/minute was found between CBF measurements performed with and without triggering (P < 0.05). For repeated triggered measurements, the coefficient of variation (CV) was 7.1%, and for nontriggered measurements 10.3%. For repeated measurements with repositioning, the CV was 7.1% with and 11.2% without triggering. Repeated measurements at different occasions showed a CV of 8.8%. Comparing measured with real flow in the phantom, the triggered differed 4.9% and the nontriggered 8.3%.ConclusionThe findings of this study demonstrate that pcMRI is a reliable method to measure total CBF in terms of both accuracy and reproducibility. J. Magn. Reson. Imaging 2002;16:1-5.
机译:目的通过相衬磁共振成像(pcMRI)评估总脑血流量(CBF)测量的可重复性。材料与方法我们在15位有或没有心脏触发,有或没有复位的健康志愿者中重复进行了总CBF测量。在八位志愿者中,分别在两个不同的场合进行了测量。此外,还进行了幻影中的流量测量以验证MR测量结果。结果在进行和不进行触发的CBF测量之间发现差异为40.4 ml /分钟(P <0.05)。对于重复触发的测量,变异系数(CV)为7.1%,而对于非触发测量,变异系数(CV)为10.3%。对于重新定位的重复测量,有触发时CV为7.1%,无触发时CV为11.2%。在不同情况下重复测量的CV为8.8%。与人体模型中的实际流量进行比较,触发的差异为4.9%,未触发的差异为8.3%。结论这项研究的结果表明,pcMRI是测量总CBF的可靠方法,无论是准确性还是可重复性。 J.Magn。雷森成像2002; 16:1-5。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号