Respiratory-triggered MRCP applying parallel acquisition techniques.

机译:应用并行采集技术的呼吸触发式MRCP。

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: To evaluate the influence of parallel imaging on the image quality of respiratory triggered magnetic resonance cholangiopancreatography (MRCP). MATERIALS AND METHODS: A total of 30 consecutive patients underwent MRCP applying a respiratory triggered T2-weighted (T2w) turbo spin-echo (TSE) sequence without and with parallel imaging (acceleration factor of 2). Acquisition times of both sequences were recorded. Quantitative evaluation included measurement of a contour sharpness index of two segments of the pancreaticobiliary tree as well as calculation of the relative contrast between ductal structures and organ parenchyma at four different segments. The qualitative evaluation was performed by two independent radiologists who graded overall image quality, depiction of eight segments of the pancreaticobiliary tree, and the frequency of artifacts. RESULTS: The application of parallel imaging significantly (P<0.05) reduced the acquisition time of the respiratory triggered MRCP sequence by 37.7% (sixminutes and two seconds+/-one minute and 26 seconds vs. three minutes and 46 seconds+/-58 seconds). The quantitative and qualitative evaluation revealed no statistically significant differences between the two sequences (P>0.05). The frequency of artifacts was at the same level for both sequences as well. CONCLUSION: The application of parallel imaging for respiratory triggered MRCP significantly reduces the acquisition time without relevant influence on image quality.
机译:目的:评估并行成像对呼吸触发磁共振胰胆管造影(MRCP)图像质量的影响。材料与方法:共有30例连续患者接受了MRCP,应用了呼吸触发的T2加权(T2w)涡轮自旋回波(TSE)序列,而没有并行成像(加速度因子为2)。记录两个序列的采集时间。定量评估包括测量胰胆管树的两个节段的轮廓清晰度指数,以及计算四个不同节段的导管结构与器官实质之间的相对对比度。定性评估由两名独立的放射科医生进行,他们对整体图像质量,胰胆管树的八个部分的描绘以及伪影的频率进行了分级。结果:平行成像的应用显着(P <0.05)将呼吸触发的MRCP序列的采集时间减少了37.7%(6分钟和2秒+/- 1分钟和26秒,而3分钟和46秒+/- 58秒) 。定量和定性评估显示两个序列之间无统计学差异(P> 0.05)。对于这两个序列,伪像的频率也处于相同的水平。结论:并行成像在呼吸触发MRCP中的应用显着减少了采集时间,而对图像质量没有相关影响。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号