Combined high-resolution and real-time imaging: A technical feasibility study on coronary magnetic resonance angiography.

机译:高分辨率和实时成像相结合:冠状动脉磁共振血管造影的技术可行性研究。

获取原文
获取原文并翻译 | 示例
           

摘要

PURPOSE: To propose a new approach to combining high-resolution and real-time imaging and to show its technical feasibility on the example of coronary magnetic resonance angiography. MATERIALS AND METHODS: The insertion of fast two-dimensional (2D) acquisitions into time intervals that have not been utilized by triggered or gated 2D or three-dimensional (3D) acquisitions so far is suggested, as well as the immediate reconstruction and display of the additional data. For a technical validation of this concept, a 2D ventricular function protocol was interleaved into a cardiac-triggered and respiratory-gated 3D coronary angiography protocol. Dedicated hardware was employed to rapidly process the data originating from the former. Since the sampling of the latter was restricted to intervals with minimal motion, remaining periods of time could be used to simultaneously image the cardiac and respiratory motion. RESULTS: The technical feasibility of the proposed approach was demonstrated by successful measurements with the combined high-resolution and real-time protocol in volunteers. All examinations provided short axis views during the acquisition and angiograms of selected parts of the coronary system after its completion. CONCLUSION: The investigated concept allows high-resolution measurements to be complemented with real-time imaging functionality without affecting the scan time or image quality. In the particular application considered, an image-based patient monitoring or motion correction is enabled, indicating potential benefits of combining two very dissimilar methods of data acquisition in one measurement. J. Magn. Reson. Imaging 2002;16:584-590.
机译:目的:提出一种将高分辨率和实时成像相结合的新方法,并以冠状动脉磁共振血管造影为例展示其技术可行性。材料和方法:建议将快速二维(2D)采集插入到到目前为止尚未被触发或门控2D或三维(3D)采集使用的时间间隔中,并建议立即重建和显示附加数据。为了对该概念进行技术验证,将2D心室功能协议插入心脏触发和呼吸门控3D冠状动脉造影协议中。采用专用硬件来快速处理源自前者的数据。由于后者的采样仅限于最小运动间隔,因此可以将剩余时间段用于同时对心脏和呼吸运动进行成像。结果:结合高分辨率和实时协议在志愿者中进行的成功测量证明了该方法的技术可行性。所有检查都在采集过程中提供了短轴视图,并在完成冠状动脉系统的选定部分后进行了血管造影。结论:所研究的概念允许高分辨率测量与实时成像功能互补,而不会影响扫描时间或图像质量。在所考虑的特定应用中,启用了基于图像的患者监视或运动校正,从而表明了将两种截然不同的数据采集方法结合到一个测量中的潜在好处。 J.Magn。雷森成像2002; 16:584-590。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号