A strip-shield improves the efficiency of a solenoid coil in probes for high-field solid-state NMR of lossy biological samples

机译:带状屏蔽提高了有损生物样品的高场固态NMR探针中螺线管线圈的效率

获取原文
获取原文并翻译 | 示例
           

摘要

A strip-shield inserted between a high inductance double-tuned solenoid coil and the glass tube containing the sample improves the efficiency of probes used for high-field solid-state NMR experiments on lossy aqueous samples of proteins and other biopolymers. A strip-shield is a coil liner consisting of thin copper strips layered on a PTFE (polytetrafluoroethylene) insulator. With lossy samples, the shift in tuning frequency is smaller, the reduction in Q and RF-induced heating are all significantly reduced when the strip-shield is present. The performance of 800 MHz H-1/N-15 and H-1/C-13 double-resonance probes is demonstrated on aqueous samples of membrane proteins in phospholipid bilayers.
机译:插入高电感双调谐螺线管线圈和装有样品的玻璃管之间的带状屏蔽可提高用于蛋白质和其他生物聚合物的有损水性样品的高场固态NMR实验的探针的效率。屏蔽带是由衬在PTFE(聚四氟乙烯)绝缘体上的薄铜带组成的线圈衬套。对于有损耗的样品,当带状屏蔽罩存在时,调谐频率的偏移会更小,Q的降低和RF引起的发热都将大大降低。在磷脂双层膜蛋白的水性样品中证明了800 MHz H-1 / N-15和H-1 / C-13双共振探针的性能。