World-smallest LVAD with 27 g weight, 21 mm OD and 5 l min-1 flow with 50 mmHg pressure increase.

机译:世界上最小的LVAD,重量为27 g,外径为21 mm,流量为5 l min-1,压力增加了50 mmHg。

获取原文
获取原文并翻译 | 示例
           

摘要

To investigate the feasibility of a long-term left ventricular assist device (LVAD) placed in the aortic valve annulus, an implantable aortic valve pump (21 mm outer diameter, weighing 27 g) was developed. The device consists of a central rotor and a stator. The rotor assembly incorporates driven magnets and an impeller. The stator assembly has a motor coil with an iron core and outflow guide vanes. The device is to be implanted identically to an aortic valve replacement, occupying no additional anatomic space. The pump delivers the blood directly from left ventricle to the aortic root, like a natural ventricle, therefore causing less physiologic disturbance to the natural circulation. Neither connecting conduits nor 'bypass' circuits are necessary. The pump is designed to cycle between a peak flow and zero net flow to approximate systole and diastole. Bench testing indicates that the pump can produce a blood flow of 5 l min(-1) with 50 mmHg pressure increase at 17,500 rpm. At zero net flow rate, the pump can maintain a diastole aortic pressure against 80 mmHg at the same rotating speed.
机译:为了研究将长期左心室辅助设备(LVAD)放置在主动脉瓣环中的可行性,开发了一种植入式主动脉瓣泵(外径21 mm,重27 g)。该设备由中央转子和定子组成。转子组件包含从动磁铁和叶轮。定子组件具有带铁芯和流出导向叶片的电动机线圈。该设备将与主动脉瓣置换相同地植入,不占用额外的解剖空间。该泵将血液从左心室像自然心室一样直接从左心室传递到主动脉根部,因此对自然循环的生理干扰较小。不需要连接导管或“旁路”电路。该泵设计为在峰值流量和零净流量之间循环,以近似收缩压和舒张压。台架试验表明,该泵可在17500 rpm的压力下增加50 mmHg的压力产生5 l min(-1)的血流。在零净流量下,泵可以在相同转速下维持80 mmHg的舒张主动脉压。

著录项

相似文献

  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号