Circular external fixators, and in particular hexapod circular external fixators, are increasingly being used in orthopaedic traumatology and limb reconstruction. Their ability to provide three dimensional stability throughout the treatment period make these external fixators indispensable in the management of complex reconstructive work. This study has identified a unique mode of failure with the use of Taylor Spatial Frame Fast-Fx struts and termed this complication catastrophic strut collapse. In this article a lab model is used to demonstrate the consequences of catastrophic strut collapse and warn users of the importance of using the ID bands and locking nuts to prevent inadvertent fragment displacement.
机译:圆形外固定器,尤其是六脚形圆形外固定器,越来越多地用于整形外科创伤学和四肢重建中。它们在整个治疗过程中提供三维稳定性的能力使得这些外固定器在管理复杂的重建工作中必不可少。这项研究通过使用泰勒空间框架Fast-Fx撑杆确定了一种独特的失效模式,并将其称为这种复杂的灾难性撑杆坍塌。在本文中,将使用一个实验室模型来演示灾难性支杆坍塌的后果,并警告用户使用ID带和锁紧螺母以防止意外碎片移位的重要性。