首页> 外文期刊>Journal of medical engineering & technology >Continuous amperometric monitoring of subcutaneous oxygen in rabbit by telemetry.
【24h】

Continuous amperometric monitoring of subcutaneous oxygen in rabbit by telemetry.

机译:遥测法连续监测兔皮下氧气。

获取原文
获取原文并翻译 | 示例
           

摘要

The primary objective was to develop a telemetric oxygen sensor designed for implantation in subcutaneous tissue. In the sensor, the platinum indicating electrode was covered with polythene to retard oxygen entry and avoid interference. Given the potential for large indicating electrodes to deplete oxygen in vivo, some sensors were made with very small indicating electrodes. Sensors were tested in the anesthetized rabbit at various oxygen concentrations. Animals were studied 1 and 5 days after implementation. The sensor could detect changes in oxygen in vivo. Measurements were similar at 1 and 5 days and sensors with different sized electrodes yielded similar results. This telemetric implantable amperometric sensor can detect variations in tissue oxygenation and may have applications in cardiopulmonary physiology, continuous glucose monitoring and other related fields.
机译:主要目的是开发一种用于皮下组织植入的遥测氧气传感器。在传感器中,铂指示电极上覆盖有聚乙烯,以阻止氧气进入并避免干扰。考虑到较大的指示电极有可能在体内消耗氧气,因此一些传感器的电极很小。在麻醉的兔子中以各种氧气浓度测试传感器。实施后1天和5天对动物进行研究。该传感器可以检测体内氧气的变化。第1天和第5天的测量结果相似,电极尺寸不同的传感器也得出了相似的结果。这种遥测式植入式安培传感器可以检测组织氧合的变化,并可以在心肺生理,连续血糖监测和其他相关领域中应用。