OBJECTIVES: To establish the reliability, accuracy and concurrent validity of electrogoniometers for measuring extreme lower extremity movements. STUDY DESIGN: Reliability, accuracy and validity study. Methods: This project employed two experiments. Investigation I determined electrogoniometer accuracy compared to a digital protractor. Investigation II compared electrogoniometers to a criterion measure, motion analysis, in sagittal plane dance movements in 17 dancers. Intraclass correlation coefficients (ICC) and standard error of measurement (SEM) were calculated for hip, knee, and ankle sagittal plane angular displacements. RESULTS: In Investigation I, electrogoniometer to protractor correlations were high (r > or = 0.998, SEM < or =3.65 degrees ). In Investigation II, instrument and intra-rater reliability correlations were high (r > or = 0.983 and r > or = 0.972, SEM < or = 3.49 degrees and < or =4.48 degrees ), as were concurrent validity correlations (r > or = 0.949, SEM < or =6.80 degrees ) to motion analysis. CONCLUSIONS: This error may be acceptable for lower extremity studies with extreme movements when motion capture is unavailable.
机译:目的:建立用于测量极端下肢运动的电动测角仪的可靠性,准确性和同时有效性。研究设计:可靠性,准确性和有效性研究。方法:该项目进行了两个实验。调查我确定了与数字量角器相比的电子测角计精度。研究二在17位舞者的矢状平面舞动作中将电子测角仪与标准测量,运动分析进行了比较。计算髋,膝和踝矢状平面角位移的类内相关系数(ICC)和标准测量误差(SEM)。结果:在研究I中,电测角仪与量角器的相关性很高(r>或= 0.998,SEM <或= 3.65度)。在调查II中,仪器和评估者内部的可靠性相关性很高(r>或= 0.983和r>或= 0.972,SEM <或= 3.49度和<或= 4.48度),同时有效效度相关性(r>或= 0.949,SEM <或= 6.80度)进行运动分析。结论:在无法进行运动捕捉的情况下,对于下肢运动剧烈运动的下肢研究,该误差是可以接受的。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号