Precision and repeatability of the Optotrak 3020 motion measurement system.

机译:Optotrak 3020运动测量系统的精度和可重复性。

获取原文
获取原文并翻译 | 示例
           

摘要

Several motion analysis systems are used by researchers to quantify human motion and to perform accurate surgical procedures. The Optotrak 3020 is one of these systems and despite its widespread use there is not any published information on its precision and repeatability. We used a repeated measures design study to evaluate the precision and repeatability of the Optotrak 3020 by measuring distance and angle in three sessions, four distances and three conditions (motion, static vertical, and static tilted). Precision and repeatability were found to be excellent for both angle and distance although they decreased with increasing distance from the sensors and with tilt from the plane of the sensors. Motion did not have a significant effect on the precision of the measurements. In conclusion, the measurement error of the Optotrak is minimal. Further studies are needed to evaluate its precision and repeatability under human motion conditions.
机译:研究人员使用了几种运动分析系统来量化人体运动并执行准确的手术程序。 Optotrak 3020是这些系统之一,尽管已得到广泛使用,但尚未发布有关其精度和可重复性的任何信息。我们通过重复测量设计研究,通过在三个阶段,四个距离和三个条件(运动,静态垂直和静态倾斜)中测量距离和角度来评估Optotrak 3020的精度和可重复性。发现精度和可重复性对于角度和距离都是极好的,尽管它们随着与传感器的距离增加和与传感器平面的倾斜而降低。运动对测量精度没有显着影响。总之,Optotrak的测量误差很小。需要进一步研究以评估其在人体运动条件下的精度和可重复性。