Periodical gait asymmetry assessment using real-time wireless gyroscopes gait monitoring system

机译:使用实时无线陀螺仪步态监测系统进行周期性步态不对称评估

获取原文
获取原文并翻译 | 示例

摘要

A real-time gait monitoring system that incorporates an immediate and periodical assessment of gait asymmetry is described. This system was designed for gait analysis and rehabilitation of patients with pathologic gait. It employs wireless gyroscopes to measure the angular rate of the thigh and shank in real time. Cross-correlation of the lower extremity (Ccnorm), and normalized Symmetry Index (SInorm) are implemented as new approaches to periodically determine the gait asymmetry in each gait cycle. Cc norm evaluates the signal patterns measured by wireless gyroscopes in each gait cycle. SInorm determines the movement differences between the left and right limb. An experimental study was conducted to examine the viability of these methods. Artificial asymmetrical gait was simulated by placing a load on one side of the limbs. Results showed that there were significant differences between the normal gait and asymmetrical gait (p0.01). They also indicated that the system worked well in periodically assessing the gait asymmetry.
机译:描述了一种实时步态监测系统,该系统结合了对步态不对称性的即时和定期评估。该系统设计用于步态分析和病理性步态患者的康复。它使用无线陀螺仪实时测量大腿和小腿的角速度。下肢(Ccnorm)和归一化对称指数(SInorm)的互相关被实现为定期确定每个步态周期中步态不对称性的新方法。 Cc规范评估每个步态周期中由无线陀螺仪测量的信号模式。 SInorm确定左,右肢之间的运动差异。进行了实验研究以检查这些方法的可行性。通过将负载施加在四肢的一侧来模拟人工不对称步态。结果表明,正常步态和不对称步态之间存在显着差异(p <0.01)。他们还指出,该系统在定期评估步态不对称性方面表现良好。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号