X-ray Detection Performance of 100-pixel Superconducting Tunnel Junction Array Detector in the Soft X-ray Region

机译:100像素超导隧道结阵列探测器在软X射线区域的X射线检测性能

获取原文
获取原文并翻译 | 示例
       

摘要

The overall performance of a 100-pixel superconducting tunnel junction (STJ) array detector in the soft X-ray region was determined by measuring fluorescent X-rays from light elements. The yield of working STJs in the fabricated detector was 90percent, and the leakage current for these 90 STJs was 11.6 + - 2.5 nA. The STJs had an intrinsic energy resolution for O-K(alpha) X-rays (524.9 eV) of 10.2 + - 1.7 eV (FWHM). The deviation from energy linearity was -0.042 + - 0.45percent, 0.793 + - 0.91percent, and 0.135 + - 1.31percent for O-K(alpha) (524.9 eV), N-K(alpha) (392.4 eV) and B-K(alpha) (183.3 eV) X-rays, respectively. Most of the STJs exhibited uniform X-ray detection characteristics. In addition, the detector could discriminate the N-K(alpha) peak in the fluorescent X-ray spectrum from n-type SiC doped with 300 ppm nitrogen.
机译:通过测量来自发光元件的荧光X射线,可以确定100像素超导隧道结(STJ)阵列探测器在软X射线区域中的整体性能。所制造的检测器中工作的STJ的产率为90%,这90个STJ的泄漏电流为11.6 +-2.5 nA。 STJ对O-KαX射线(524.9 eV)的本征能量分辨率为10.2 +-1.7 eV(FWHM)。对于OKα(524.9 eV),NKα(392.4 eV)和BKα(183.3),与能量线性的偏差为-0.042 +-0.45%,0.793 +-0.91%和0.135 +-1.31% eV)分别是X射线。大多数STJ显示出均匀的X射线检测特性。另外,检测器可以从掺杂有300 ppm氮的n型SiC中区分出荧光X射线光谱中的N-Kα峰。