首页> 外文期刊>Journal of mass spectrometry: JMS >Simple correction improving long-term reproducibility of HPLC-MS
【24h】

Simple correction improving long-term reproducibility of HPLC-MS

机译:简单校正即可提高HPLC-MS的长期重现性

获取原文
获取原文并翻译 | 示例
获取外文期刊封面目录资料

摘要

Chromatographic peak areas in long series of high-performance liquid chromatography-MS experiments often vary, which decrease reproducibility and may cause bias in the results. It was found that the sensitivity of various components change differently; in our case, variability is in the order of 20-40%, and it is most likely due to changing conditions in electrospray ionization (ESI). The most often used peak area correction methods do not take this effect into account. The change in peak areas can be well described by a polynomial function; we found that a fourth-order polynomial is most often suitable. We suggest a simple correction algorithm based on polynomial fitting. When the experiments were inherently well reproducible, this correction improved reproducibility from 12% to 3% (on average for various components). When random errors were larger, this improvement was less significant (15% to 12% in nano-ESI) but nevertheless essential in order to avoid possible bias in the results. Copyright (c) 2015 John Wiley & Sons, Ltd.
机译:长时间进行的高性能液相色谱-MS实验中的色谱峰面积经常变化,这会降低重现性并可能导致结果偏差。结果发现,各种成分的灵敏度变化不同。在我们的案例中,变异性大约在20%至40%之间,这很可能是由于电喷雾电离(ESI)条件的变化所致。最常用的峰面积校正方法没有考虑到这种影响。峰面积的变化可以通过多项式函数很好地描述。我们发现四阶多项式最适合。我们建议一种基于多项式拟合的简单校正算法。当实验固有地具有良好的可重复性时,这种校正将可重复性从12%提高到3%(各种组分的平均值)。当随机误差较大时,这种改进的意义就不那么明显了(在nano-ESI中为15%到12%),但是为了避免结果可能出现偏差,这种改进必不可少。版权所有(c)2015 John Wiley&Sons,Ltd.

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号