Direct and indirect catheter-to-catheter carbonization caused by high power radiofrequency delivery via a large tip ablation electrode.

机译:经由大尖端消融电极的高功率射频输送导致导管间直接碳化和间接碳化。

获取原文
获取原文并翻译 | 示例

摘要

We report two cases of directly and indirectly induced catheter carbonizations by radiofrequency application via a large tip ablation catheter. We assume that during a large tip high power delivery of more than 50 watts, an adjacent, smaller-sized, low resistance electrode produces a local increase in the intensity of the current field that is sufficient to elevate the tissue temperature above 100 degrees C. Due to the potential risk of embolism, this may have an impact on ablation procedures in the left atrium and ventricle using similar mapping configurations.
机译:我们报告了通过大尖端消融导管通过射频应用直接和间接诱发导管碳化的两种情况。我们假设在大于50瓦的大尖端大功率输送过程中,相邻的较小尺寸的低电阻电极在电流场强度上产生局部升高,足以使组织温度升高至100摄氏度以上。由于存在栓塞的潜在风险,使用类似的映射配置,这可能会对左心房和心室的消融程序产生影响。