Effect of oversized partial postdilatation on coatings of contemporary durable polymer-based drug-eluting stents: a scanning electron microscopy study.

机译:超大型部分后扩张对当代耐用的基于聚合物的药物洗脱支架涂层的影响:扫描电子显微镜研究。

获取原文
获取原文并翻译 | 示例
           

摘要

BACKGROUND: Oversized postdilation of drug-eluting stents (DES) is often performed to avoid stent malapposition. In stents implanted in long lesion or major bifurcations, extremely oversized partial postdilation may be required, which exposes DES coating to extreme forces. This study aims to assess shape and incidence of coating irregularities on durable polymer-based DES following extremely oversized partial postdilatation. METHODS: Fifteen DES samples (3 3.5 mm stents of Cypher Select plus [Cordis Europa, Roden, the Netherlands], Taxus Liberte[Boston Scientific Corp., Natick, MA, USA], Endeavor Sprint [Medtronic Vascular, Santa Rosa, CA, USA], Endeavor Resolute [Medtronic Vascular, Santa Rosa, CA, USA], and Xience V [Abbott Vascular, Santa Clara, CA, USA]) were deployed in sterile water (37 degrees C) at 14 atm, followed by a proximal postdilation with noncompliant 5.0-mm balloons at 18 atm. Stents were then examined with scanning electron microscopy. RESULTS: Thorough examination of a total of 660 scanning electron microscopic images demonstrated that shape and incidence of coating irregularities in the postdilated and/or transitional DES regions differed only mildly from the nonpostdilated regions. Cypher Select plus showed more peeling without bare metal aspect in the postdilated and transitional regions, and cracks were wider (P < 0.001) in the postdilated and transitional regions; in Taxus Liberte one additional irregularity (torn webbing) and more wrinkles were observed (P < 0.05 for both); in Endeavor Resolute wider cracks were found in the extremely postdilated region only (P < 0.001). Endeavor Sprint and Xience V showed no differences in shape or incidence of coating irregularities between oversized and nonoversized stent regions. CONCLUSIONS: Bench side assessment of five contemporary durable polymer-based DES with scanning electron microscopy suggests that even very aggressive stent postdilatation results in no more than mild differences in coating irregularities between postdilated and nonpostdilated stent regions.
机译:背景:药物洗脱支架(DES)的超大后扩张通常是为了避免支架贴壁不良。在植入了长病变或主要分叉的支架中,可能需要使用尺寸过大的局部后扩张,这会使DES涂层承受极大的力。这项研究的目的是评估部分过大的局部扩张后,基于耐用聚合物的DES的涂层不规则形状和发生率。方法:15个DES样品(3个3.5毫米Cypher Select plus支架[Cordis Europa,Roden,荷兰],Taxus Liberte [波士顿科学公司,内蒂克,马萨诸塞州,美国),Endeavour Sprint [Medtronic Vascular,Santa Rosa,CA,美国],Endeavor Resolute [Medtronic Vascular,美国加利福尼亚州圣塔罗莎]和Xience V [Abbott Vascular,美国加利福尼亚州圣塔克拉拉]]分别在14个大气压的无菌水中(37摄氏度)部署,然后在近端在18个大气压下使用不兼容的5.0毫米气球进行后扩张。然后用扫描电子显微镜检查支架。结果:对总共660幅扫描电子显微镜图像的彻底检查表明,后扩张和/或过渡DES区域的涂层不规则形状和发生率与未扩张后的区域仅稍有不同。 Cypher Select plus在后扩散区和过渡区显示出更多的剥离现象,而没有裸露的金属外观,并且在后膨胀区和过渡区的裂纹变宽了(P <0.001)。在南方红豆杉中,又发现一种不规则现象(撕裂的织带)和更多的皱纹(两者均P <0.05);在“奋进号”中,仅在极度后扩张的区域发现了较宽的裂纹(P <0.001)。奋进号Sprint和Xience V在大型和非大型支架区域之间的形状或涂层不规则发生率均无差异。结论:用扫描电子显微镜对五个当代耐用的基于聚合物的DES进行了台面评估,结果表明,即使非常强力的支架后扩张,其后扩张和非后扩张支架区域之间的涂层不规则性差异也不会超过温和的差异。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号