首页> 外文期刊>Journal of interventional cardiac electrophysiology: an international journal of arrhythmias and pacing >'Pseudo' Faraday cage: a solution for telemetry link interaction between a left ventricular assist device and an implantable cardioverter defibrillator.
【24h】

'Pseudo' Faraday cage: a solution for telemetry link interaction between a left ventricular assist device and an implantable cardioverter defibrillator.

机译:“伪”法拉第笼:一种用于左心室辅助设备与可植入式心脏复律除颤器之间遥测链路相互作用的解决方案。

获取原文
获取原文并翻译 | 示例
           

摘要

Patients implanted with left ventricular assist devices (LVAD) may have implantable cardioverter defibrillators (ICD) implanted for sudden cardiac death prevention. This opens the possibility of device-device communication interactions and thus interferences. We present a case of such interaction that led to ICD communication failure following the activation of an LVAD. In this paper, we describe a practical solution to circumvent the communication interference and review the communication links of ICDs and possible mechanisms of ICD-LVAD interactions.
机译:植入左心室辅助设备(LVAD)的患者可能植入了可植入的心脏复律除颤器(ICD),以预防心脏猝死。这打开了设备与设备之间的通信交互以及由此产生干扰的可能性。我们介绍了这样一种交互情况,导致在激活LVAD之后导致ICD通信失败。在本文中,我们描述了一种避免通信干扰的实用解决方案,并回顾了ICD的通信链接以及ICD-LVAD交互作用的可能机制。