首页> 外文期刊>Journal of interventional cardiac electrophysiology: an international journal of arrhythmias and pacing >Inappropriate Single Chamber ICD Discharges due to Supraventricular Tachycardia with High Degree Atrioventricular Block.
【24h】

Inappropriate Single Chamber ICD Discharges due to Supraventricular Tachycardia with High Degree Atrioventricular Block.

机译:室上性心动过速伴高度房室传导阻滞导致单室ICD放电不当。

获取原文
获取原文并翻译 | 示例
           

摘要

Supraventricular tachycardia with rapid ventricular response is well recognized as the more frequent cause of single chamber ICD inappropriate therapies.We report here a 18-year-old-woman with surgically corrected transposition of the great arteries who received repetitive inappropriate discharges from an ICD implanted for ventricular tachycardia. Rapid atrial activity during episodes of supraventricular tachycardia with high degree atrioventricular block was oversensed as ventricular fibrillation by a single chamber ICD causing repetitive painful discharges. Pharmacological treatment of the supraventricular tachycardia solved the problem.
机译:室上性心动过速伴快速心室反应是公认的单腔室ICD不适当治疗的较常见原因。我们报道了一名18岁女性,经手术矫正了大动脉移位,该患者接受了从植入ICD的反复不适当放电室性心动过速。单室ICD引起室上性心动过速伴高度房室传导阻滞发作时快速的心房活动被过度感知为室颤,引起重复性疼痛放电。室上性心动过速的药物治疗解决了这个问题。