Accuracy of the OMRON M4 automatic blood pressure measuring device.

机译:OMRON M4自动血压测量设备的准确性。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

Despite widespread use of the automatic blood pressure (BP) measuring device OMRON M4, there is little formal validation on its accuracy. A study was conducted to assess the accuracy of the OMRON M4 compared with the true indirect BP measured by mercury sphygmomanometer (MS). A rapid method for the evaluation of automatic blood pressure measurement devices (READ) was applied for this study. The READ is based on numerous BP measurements at rest and during a standardised postural challenge in a small number of subjects who exhibit a wide range of BPs. Blood pressure measurements were done in supine position for 10 min followed by head-up tilt for 30 min and again supine for 10 min. The automatic device (AU) and the MS were connected to one arm-cuff by means of a T connector. A stethoscope with dual sets of ear-pieces was used for duplicate MS measurements (MS1 and MS2). The MS1, MS2 and AU measurements were taken simultaneously in a blinded manner. Three units of the automatic instrument were evaluated. An average of 111 measurements per unit were performed, every BP category being present in >/=15 MS measurements. The differences between MS1 and MS2 measurements (DeltaBP: MS1-MS2) were utilized to assess the consistency of true indirect BP and the differences between AU and MS measurements (DeltaBP:AU-MS2) were utilized to assess the accuracy of the AU. The following characteristics of the OMRON M4 were assessed: (1) grade of accuracy, (2) aberration pattern, (3) consistency of the aberration-pattern, and (4) correlation between levels of BP and DeltaBP: AU-MS. For MS paired readings, 92-100% of systolic and 99-100% of diastolic readings fell within 5 mm Hg difference range, that is consistent with a British Hypertension Society grade A of both. For AU compared to MS2, 29-64% of systolic and 73-94% of diastolic readings fell within 5 mm Hg and 49-86% of systolic and 86-99% of diastolic readings fell within 10 mm Hg and the devices qualified C, D and C, respectively. All devices exhibited irregular and inconsistent aberration patterns, making the design of correction formulas impractical. In conclusion, the OMRON M4 device did not meet the requirements of the British Hypertension Society and, therefore, cannot be recommended for clinical use.
机译:尽管自动血压(BP)测量设备OMRON M4被广泛使用,但对其准确性的正式验证很少。进行了一项研究,以评估OMRON M4与通过水银血压计(MS)测量的真实间接BP的准确性。本研究采用了一种评估自动血压测量装置(READ)的快速方法。 READ是基于在静止状态下以及在标准化姿势挑战期间,对表现出广泛BP的少数受试者的大量BP测量得出的。血压测量是在仰卧位进行10分钟,然后抬起头倾斜30分钟,然后再次仰卧10分钟。自动装置(AU)和MS通过T型连接器连接到一个袖套上。使用带有两组听筒的听诊器进行重复的MS测量(MS1和MS2)。 MS1,MS2和AU测量以盲法同时进行。对自动仪器的三个单元进行了评估。平均执行每单位111次测量,每个BP类别均出现在> / = 15 MS测量中。 MS1和MS2测量值之间的差异(DeltaBP:MS1-MS2)用于评估真实间接BP的一致性,而AU和MS测量值之间的差异(DeltaBP:AU-MS2)用于评估AU的准确性。评估了OMRON M4的以下特征:(1)精度等级,(2)像差模式,(3)像差模式的一致性,以及(4)BP和DeltaBP水平之间的相关性:AU-MS。对于MS配对读数,收缩压读数的92-100%和舒张压读数的99-100%落在5 mm Hg的差异范围内,这与英国高血压协会的A级都一致。与MS2相比,AU的收缩压读数的29-64%和舒张压读数的73-94%落在5 mm Hg之内,收缩压读数的49-86%和舒张压读数的86-99%落在10 mm Hg之内,并且设备合格C ,D和C。所有设备均显示出不规则且不一致的像差模式,因此校正公式的设计不切实际。总之,OMRON M4设备不符合英国高血压协会的要求,因此不建议用于临床。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号