首页> 外文期刊>Journal of human hypertension >Measurement of blood pressure using the auscultatory and oscillometric methods in the same cuff deflation: validation and field trial of the A&D TM2421 monitor.
【24h】

Measurement of blood pressure using the auscultatory and oscillometric methods in the same cuff deflation: validation and field trial of the A&D TM2421 monitor.

机译:在相同的袖带放气中使用听诊法和示波法测量血压:A&D TM2421监护仪的验证和现场试验。

获取原文
获取原文并翻译 | 示例
   

获取外文期刊封面封底 >>

       

摘要

We have evaluated under laboratory validation conditions and in an extensive field trial the behaviour of an ambulatory monitoring device that is capable of recording both by the Korotkoff-sound and oscillometric methods in a single cuff deflation (TM2421: A&D Co, Tokyo, Japan). The effects of subject age and blood pressure (BP) level on the accuracy and field reliability of the two methods implemented in the device have been determined. In the validation phase, automatic BP measurements were compared with readings by two trained observers in 96 subjects, and the results compared with the AAMI criteria for automatic BP monitors. In the field trial phase, the performances of Korotkoff-sound and oscillometric methods over a 24-h period of ambulatory BP monitoring were compared in 515 subjects, with analysis of the agreement between the two methods in patients where both provided satisfactory recordings. In the validation phase, the Korotkoff-sound method gave satisfactory results for both systolic and diastolic BP, but the oscillometric technique narrowly failed to meet the AAMI criteria for the measurement of either systolic or diastolic BP. In the field trial, the K-sound method failed to record BP accurately in 12% of subjects whereas the oscillometric method was successful in all of these. Where both methods provided technically adequate records, agreement between mean values for each method was close. In 18% of patients, the availability of the oscillometric measurement as a 'back-up' method for the K-sound method significantly improved the number of available measurements in the monitoring period, which should result in improved accuracy and reproducibility of the ambulatory mean values.
机译:我们已经在实验室验证条件下进行了评估,并在广泛的现场试验中评估了一种能够通过Korotkoff声音和示波法同时记录一次袖带放气的动态监测设备的性能(TM2421:A&D Co,Tokyo,Japan)。已经确定了受试者年龄和血压(BP)水平对在设备中实施的两种方法的准确性和现场可靠性的影响。在验证阶段,将自动BP测量与两名受过训练的观察员对96位受试者的读数进行比较,并将结果与​​自动BP监护仪的AAMI标准进行比较。在现场试验阶段,比较了515名受试者在24小时动态BP监测中使用Korotkoff声音法和示波法的性能,并分析了两种方法在患者中的一致性,并提供了令人满意的记录。在验证阶段,Korotkoff声音法对收缩压和舒张压都给出了令人满意的结果,但示波法技术几乎无法满足AAMI的收缩压或舒张压测量标准。在现场试验中,K-声音法未能准确记录12%受试者的BP,而示波法在所有这些受试者中均成功。如果两种方法都提供了技术上足够的记录,则每种方法的平均值之间的一致性很差。在18%的患者中,示波法测量作为K声法的“备用”方法的可用性显着提高了监测期间可用测量的数量,这应导致动态平均值的准确性和可重复性提高价值观。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号