Peg Restrained Intrinsic Muscle Evaluator (PRIME): development, reliability, and normative values of a device to quantify intrinsic hand muscle strength in children.

机译:钉子约束的内在肌肉评估器(PRIME):一种用于量化儿童内在手部肌肉力量的设备的开发,可靠性和规范性值。

获取原文
获取原文并翻译 | 示例
       

摘要

PURPOSE: There is a need for objective measures of pediatric intrinsic hand muscle strength as the current standard, the manual muscle test, lacks sensitivity to clinically important changes in muscle strength. We report the development, reliability, and normative values of the Peg Restrained Intrinsic Muscle Evaluator (PRIME), a device that quantifies intrinsic hand muscle strength. METHODS: Typically developing children, ages 4 to 16 years (n = 119), established normative values of intrinsic strength for thumb palmar abduction, thumb opposition, and index and small finger abduction. A subset of 30 children (15 boys, 15 girls), ages 7 to 16 years, determined inter-rater and intra-rater reliability. We calculated mean, standard deviation, intraclass correlation coefficients, and smallest detectable differences. RESULTS: Normative results indicate that gender and age were significant predictors of strength. Although the dominant hand generated higher strength measurements on average, differences were not statistically significant. Mean index and small finger abduction strength was significantly lower than thumb abduction and opposition in both genders. Intraclass correlation coefficients ranged from 0.85 to 0.94 for inter-rater reliability and 0.88 to 0.98 for intra-rater reliability. Bland-Altman plots showed an even distribution across the zero line. CONCLUSIONS: The PRIME device is a reliable tool for the quantification of intrinsic hand muscle strength in children. Age-specific and gender-specific normative values in typically developing children can serve as a future resource for clinicians treating pediatric hand or neuromuscular conditions.
机译:目的:有必要客观地测量儿科固有手部肌肉力量,将其作为当前的标准,即手动肌肉测试,缺乏对临床上重要的肌肉力量变化的敏感性。我们报告了钉住约束的固有肌肉评估器(PRIME)的发展,可靠性和规范价值,PRIME是一种量化手部固有肌肉力量的设备。方法:通常为4至16岁(n = 119)的发育中儿童,确立了拇指手掌外展,拇指对立,食指和小指外展的内在力量的规范值。由30名儿童(15名男孩,15名女孩)组成的一个子集,年龄7至16岁,确定了评估者之间和评估者内部的可靠性。我们计算了均值,标准差,类内相关系数和最小可检测差异。结果:规范结果表明性别和年龄是力量的重要指标。尽管优势手平均产生了更高的强度测量值,但差异在统计学上并不显着。男女的平均指数和小指外展强度均明显低于拇指外展和对立。对于评估者之间的可靠性,类别内相关系数的范围为0.85至0.94,对于评估者内部的可靠性,类别内相关系数的范围为0.88至0.98。 Bland-Altman图显示了零线上的均匀分布。结论:PRIME设备是定量儿童内在手部肌肉力量的可靠工具。在典型发育中的儿童中,针对特定年龄和性别的规范性价值观可以作为临床医生治疗小儿手部或神经肌肉疾病的未来资源。

著录项

相似文献

  • 外文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号