Purpose: Volar plating for distal radius fractures has caused extensor tendon ruptures resulting from dorsal screw prominence. This study was designed to determine the biomechanical impact of placing unicortical distal locking screws and pegs in an extra-articular fracture model. Methods: We applied volar-locking distal radius plates to 30 osteoporotic distal radius models. We divided radiuses into 5 groups based on distal locking fixation: bicortical locked screws, 3 lengths of unicortical locked screws (abutting the dorsal cortex [full length], 75% length, and 50% length to dorsal cortex), and unicortical locked pegs. Distal radius osteotomy simulated a dorsally comminuted, extra-articular fracture. We determined each construct's stiffness under physiologic loads (axial compression, dorsal bending, and volar bending) before and after 1,000 cycles of axial conditioning and before axial loading to failure (2 mm of displacement) and subsequent catastrophic failure. Results: Cyclic conditioning did not alter the constructs' stiffness. Stiffness to volar bending and dorsal bending forces were similar between groups. Final stiffness under axial load was statistically equivalent for all groups: bicortical screws (230 N/mm), full-length unicortical screws (227 N/mm), 75% length unicortical screws (226 N/mm), 50% length unicortical screws (187 N/mm), and unicortical pegs (226 N/mm). Force at 2-mm displacement was significantly less for 50% length unicortical screws (311 N) compared with bicortical screws (460 N), full-length unicortical screws (464 N), 75% length unicortical screws (400 N), and unicortical pegs (356 N). Force to catastrophic fracture was statistically equivalent between groups, but mean values for pegs (749 N) and 50% length unicortical (702 N) screws were 16% to 21% less than means for bicortical (892 N), full-length unicortical (860 N), and 75% length (894 N) unicortical constructs. Conclusions: Locked unicortical distal screws of at least 75% length produce construct stiffness similar to bicortical fixation. Unicortical distal fixation for extra-articular distal radius fractures should be entertained to avoid extensor tendon injury because this technique does not appear to compromise initial fixation. Clinical relevance: Using unicortical fixation during volar distal radius plating may protect extensor tendons without compromising fixation.
机译:目的:手掌钢板用于for骨远端骨折已导致由于背侧螺钉突出而引起的伸肌腱破裂。本研究旨在确定在关节外骨折模型中放置单皮质远端锁定螺钉和钉的生物力学影响。方法:我们将掌侧锁定distal骨远端板应用于30个骨质疏松性distal骨远端模型。我们根据远侧锁定固定将半径分为5组:双皮质锁定螺钉,3种长度的单皮质锁定螺钉(邻接背皮质[全长],75%长度和50%背侧皮质长度)和单皮质锁定钉。 radius骨远端截骨模拟了背粉碎性关节外骨折。我们确定了在生理条件(轴向压缩,背侧弯曲和掌侧弯曲)下每个结构在1,000个轴向调节循环之前和之后,轴向施加破坏(2 mm位移)和随后的灾难性破坏之前的刚度。结果:循环条件不会改变结构的刚度。两组之间的手掌弯曲刚度和背侧弯曲力相似。所有组在轴向载荷下的最终刚度在统计上均相等:双皮质螺钉(230 N / mm),全长单皮质螺钉(227 N / mm),75%长度单皮质螺钉(226 N / mm),50%长度单皮质螺钉(187 N / mm)和单皮质钉(226 N / mm)。 50%长的单皮质螺钉(311 N)在2mm位移下的力明显小于双皮质螺钉(460 N),全长单皮质螺钉(464 N),75%长的单皮质螺钉(400 N)和单皮质钉(356 N)。两组之间发生灾难性骨折的力在统计学上是相等的,但钉(749 N)和50%长的单皮质(702 N)螺钉的平均值比双皮质(892 N),全长单皮质(平均值)低16%至21%。 860 N)和75%长(894 N)的单皮质结构。结论:锁定的至少70%长的单皮质远端螺钉产生的结构刚度类似于双皮质固定。应考虑对关节外远端radius骨骨折进行单皮质远端固定,以避免伸肌腱损伤,因为这种技术似乎并不影响初始固定。临床意义:在vol骨远端distal骨钢板固定术中使用单皮质固定可以保护伸肌腱而不会影响固定。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号