Determination of blood leukocyte concentration with constant volume acquisition on a flow cytometer is comparable to individualized single platform testing with beads as internal reference standard.

机译:在流式细胞仪上以恒定体积采集血白细胞浓度的测定与以珠子作为内部参考标准的个性化单平台测试相当。

获取原文
获取原文并翻译 | 示例
           

摘要

Flow cytometers have a constant flow rate. This enables flow cytometers to measure leukocyte concentrations in a determined volume by acquiring data at a fixed rate over a fixed time and is called constant volume acquisition (CVA). The volume aspirated by a FACS Calibur flow cytometer in 4 min at a high rate has a median of 163 microl (IQR 156-170) with TruCount tubes. Leukocyte concentrations of 26 healthy volunteers were measured twice on up to four occasions with a Burker-Turk chamber, by single platform technology (SPT) with TruCount tubes and on the same data set using CVA. Total leukocyte concentrations determined by CVA correlated better with measurements in a Burker-Turk (BT) chamber than with SPT. Concentrations determined with CVA were 1.86% higher than with BT whereas SPT data were 5.35% higher than BT (p<0.001), and 3.36% higher than CVA (p<0.001). At leukocyte concentrations <6 million/ml SPT correlated better with BT than CVA. The SPT measurement may be more variable because it depends onmeasurement of the number of beads aliquoted, the number of beads and leukocytes aspirated, where both BT counting and CVA measurements only depend on the number of leukocytes counted. CVA with PanLeukoGating can be established using microscopy as a reference, and is comparable to BT chamber and SPT determination. Leukocyte concentrations can be measured with CVA on flow cytometers in research and clinical settings.
机译:流式细胞仪具有恒定的流速。这使流式细胞仪能够通过在固定时间内以固定速率采集数据来测量确定体积的白细胞浓度,这被称为恒定体积采集(CVA)。用FACS Calibur流式细胞仪在4分钟内以高速率抽吸的体积中位数为163微升(IQR 156-170),带有TruCount管。通过Burker-Turk室,使用TruCount管的单平台技术(SPT)和使用CVA在相同数据集上,对多达26次健康志愿者的白细胞浓度进行了两次多达两次的测量。通过CVA确定的总白细胞浓度与在Burker-Turk(BT)舱中进行的测量相比,与使用SPT具有更好的相关性。 CVA测定的浓度比BT高1.86%,而SPT数据比BT高5.35%(p <0.001),比CVA高3.36%(p <0.001)。在白细胞浓度<600万/ ml时,SPT与BT的相关性比CVA更好。 SPT测量可能更具可变性,因为它取决于等分珠数,吸出的珠和白细胞数的测量,其中BT计数和CVA测量均仅取决于计数的白细胞数。可以使用显微镜作为参照建立具有PanLeukoGating的CVA,并且可以与BT室和SPT测定相媲美。在研究和临床环境中,可以在流式细胞仪上用CVA测量白细胞浓度。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号