There are several instruments available to measure grip strength, but some instruments are costly, time-consuming to use, or have questionable reliability. The purpose of this study is to examine the concurrent validity of the Dexter Evaluation System with Jamar dynamometer attachment (Dexter) compared with the reference-based criterion of the Jamar adjustable hand dynamometer (Jamar) for measurement of maximal hand grip strength among normal subjects. Sixty-two subjects between the ages of 20 and 50 years, who had no history of hand, arm, shoulder, or neck injuries, were tested with the Jamar in the second handle position and, during the same visit, with the Dexter in the identical position. The Jamar was found to be highly reliable (ICC [3,1] = 0.98) and valid (ICC (2,K) = 0.99) for measuring hand grip strength. In this study, the Dexter was shown to be valid when compared to the Jamar dynamometer for measuring hand grip strength.
机译:有几种仪器可用来测量握力,但有些仪器价格昂贵,使用费时或可靠性令人怀疑。这项研究的目的是检查与Jamar可调测功机(Jamar)的基于参考的标准相比,带有Jamar测力计附件(Dexter)的德克斯特评估系统的并发有效性,该标准用于测量正常受试者的最大握力。在20至50岁之间的62名无手,手臂,肩膀或颈部受伤史的受试者中,Jamar处于第二把手位置,在同一次探访中,Dexter处于第二把手位置。相同的位置。发现Jamar高度可靠(ICC [3,1] = 0.98),并且有效(ICC(2,K)= 0.99)用于测量手握强度。在这项研究中,与Jamar测力计相比,Dexter被证明是有效的,可以测量手握强度。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号