IMPROVED DESIGN OF AUTOMATIC LUMINOMETER FOR TOTAL BACTERIA NUMBER DETECTION BASED ON ATP BIOLUMINESCENCE

机译:基于ATP生物发光的细菌总数自动检测仪的改进设计。

获取原文
获取原文并翻译 | 示例
           

摘要

An improved design of automatic luminometer for rapid detection of total bacterial number by adenosine 5'-triphosphate (ATP) bioluminescence was developed. To reduce the impact of bioluminescence absorption and scattering, an automatic optical calibration unit was embedded into the luminometer. An automatic injection unit was integrated to add reaction reagents automatically for accuracy of quantitative reagents and reaction time. The original data of bioluminescence intensity were fitted to improvethe detection range of the improved luminometer. After calibration using bacterial suspension containing known quantity of Escherichia coli and Staphylococcus aureus, the improved luminometer was used to measure more than 500 food samples. The results showed a satisfactory linear response (R = 0.858) when comparing with the results of conventional plate count method in the bacterial concentration range from 10 to 10~8 cfu/mL. The reproducibility of the luminometer was tested by 12 tests of standard E. coli and S. aureus suspension and the coefficient variation was 8.1%.
机译:开发了一种改进的自动发光计设计,可通过5'-三磷酸腺苷(ATP)生物发光快速检测细菌总数。为了减少生物发光吸收和散射的影响,将自动光学校准单元嵌入到发光计中。集成了一个自动进样单元,可自动添加反应试剂,以确保定量试剂和反应时间的准确性。拟合了生物发光强度的原始数据以改善改进的发光计的检测范围。在使用含有已知量的大肠杆菌和金黄色葡萄球菌的细菌悬浮液进行校准后,使用改进的发光计测量了500多个食品样品。与常规平板计数法的结果相比,在细菌浓度为10至10〜8 cfu / mL时,结果显示出令人满意的线性响应(R = 0.858)。通过12个标准大肠杆菌和金黄色葡萄球菌悬浮液的测试,测试了光度计的再现性,系数变化为8.1%。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号