首页> 外文期刊>Clinical Pediatrics >Neonatal infrared axillary thermometry.
【24h】

Neonatal infrared axillary thermometry.

机译:新生儿红外腋温计。

获取原文
获取原文并翻译 | 示例
           

摘要

The authors compared axillary skin temperatures (AT) measured with an infrared (IR) thermometer (Lightouch Neonate, Exergen Corp) with rectal temperatures (RT) in 16 newly born term infants under radiant warmers (RW) and in cribs. Twelve stable, growing premature infants in incubators were also studied. This new device may be useful because of safety and rapid results (1 second), but clinical accuracy is unknown. For term infants, mean (SD) RT-AT difference was 0.1 (0.48) degree C under RW and 0.25 (0.17) degree C 2 hours later in cribs. For premature infants in incubators the mean RT-AT difference was 0.09 (0.16) degree C. Axillary temperatures measured by IR thermometer approximate RT for newly born term infants in cribs and stable premature infants in incubators. For newly born term infants under RW, RT-AT differences vary more widely, limiting clinical usefulness in this setting. The device, the unique age of this population, and the RW environment may play a role.
机译:作者比较了用红外线(IR)温度计(Lightouch Neonate,Exergen Corp)测得的腋下皮肤温度(AT)与在辐射加热器(RW)下和婴儿床中的16个新生儿的直肠温度(RT)。还研究了十二个稳定,成长中的早产婴儿。由于安全性和快速的结果(1秒),此新设备可能有用,但是临床准确性尚不清楚。对于足月婴儿,在婴儿床下,平均(SD)RT-AT差异为0.1(0.48)摄氏度,两小时后为0.25(0.17)摄氏度。对于保育箱中的早产婴儿,平均RT-AT差异为0.09(0.16)摄氏度。通过IR温度计测得的腋窝温度约为婴儿床中新生婴儿和保育箱中稳定的早产儿的RT。对于RW以下的足月儿,RT-AT差异变化更大,限制了这种情况下的临床实用性。设备,该人群的独特年龄以及RW环境可能会起作用。