Background. Accurate determination and detection of fever is essential in the appropriate treatment of pediatric population. It is widely known that improper definitions of fever can cause grave and dangerous consequences in medical procedures. Infrared tympanic thermometry seems a relatively new and popular alternative for traditional measurement in the diagnosis of pediatric fever. However, its accuracy in the diagnosis of fever remains a major concern. Design. Systematic review and meta-analysis. Data Sources. Medline, Ovid, Elsevier, Google Scholar, and Cochrane library. Study Selection. Cross-sectional, prospective design. Data Extraction. Two investigators independently assessed selected studies and extracted data. Disagreements were resolved by discussion with other reviewers. Results. A total of 25 articles were included in our meta-analysis. The summary estimates revealed that the pooled sensitivity was 0.70 (95% confidence interval [CI] = 0.68-0.72), pooled specificity was 0.86 (95% CI = 0.85-0.88), and pooled diagnostic odds ratio was 47.3 (95% CI = 29.76-75.18), for the diagnosis of fever using infrared tympanic thermometry. Additionally, the area under the summary receiver operating characteristic curve was 0.94, and Q* value was 0.87. Conclusion. A total of 25 articles that encompassing 31 studies were analyzed. Based on our meta-analysis, accuracy of infrared tympanic thermometry in diagnosing fever is high. We can cautiously make conclusion that infrared tympanic thermometry should be widely used as fever of thermometer.
机译:背景。正确地确定和检测发烧对于适当治疗小儿人群至关重要。众所周知,不正确的发烧定义会在医疗程序中造成严重的危险后果。红外鼓膜测温法似乎是诊断小儿发热的传统测量的一种相对较新的流行方法。然而,其在发烧诊断中的准确性仍然是主要关注的问题。设计。系统评价和荟萃分析。数据源。 Medline,Ovid,Elsevier,Google Scholar和Cochrane库。研究选择。横断面,前瞻性设计。数据提取。两名研究人员独立评估了选定的研究并提取了数据。分歧通过与其他审阅者的讨论得以解决。结果。我们的荟萃分析总共包括25篇文章。汇总估算显示,合并的敏感性为0.70(95%置信区间[CI] = 0.68-0.72),合并的特异性为0.86(95%CI = 0.85-0.88),合并的诊断比值比为47.3(95%CI = 29.76-75.18),用于使用红外鼓膜测温法诊断发烧。此外,摘要接收器工作特性曲线下的面积为0.94,Q *值为0.87。结论。共分析了25篇文章,涵盖31项研究。根据我们的荟萃分析,红外鼓膜测温法诊断发烧的准确性很高。我们可以谨慎地得出结论,红外鼓膜温度计应广泛用作温度计发烧。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
获取原文

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号